PHYSITEMP PTM1 Operating Manual Download Page 13

13 

 

ERROR DISPLAYS AND TROUBLESHOOTING       

 

 

 

 

LCD 
display 

Additional 
LCD 
message 

Mode

Possible cause(s)
[Action(s)] 

"

ovEr

(OVER) 

 MEASURE 

   

a. Temperature 
measured > the 
instrument’s range for 
the thermocouple type 
sensor used. 
b. Probe not properly 
connected. 
[Change sensor with 
proper  temperature 
range.] 
[Check probe 
connection.] 

"

undr

(UNDER) 

 MEASURE 

 

a. Temperature 
measured < the 
instrument’s range for 
the thermocouple type 
sensor used. 
b. Probe not properly 
connected. 
[Change sensor with 
proper  temperature 
range.] 
[Check probe 
connection.] 

"

ovEr

(OVER) 

MAX

” 

MAX   

Unit recorded an over 
temperature reading 
before. 
[Clear the MAX values] 

"

undr

(UNDER) 

MIN

” 

MIN 

Unit  recorded an 
under temperature 
reading before. 
[Clear the MIN values] 

"

FUL

(FULL) 

SAVE

” 

MEASURE 

User tries to save data 
but the 50 sites are full. 
[Go to the RECALL 
mode and erase all 
data] 

EPt

” 

(EMPTY) 

RECALL

” 

RECALL 

User tried to go to the 
RECALL mode but no 
data has been saved. 
[Save at least 1 set of 

data] 

Summary of Contents for PTM1

Page 1: ...physitemp Operating Manual Portable Temperature Meter PTM1...

Page 2: ...2...

Page 3: ...3...

Page 4: ...S FUNCTION 5 MODEL PTM1 MODES 7 A MEASURE mode 7 B MAX mode 8 C MIN mode 8 D HOLD mode 8 E SETTING mode 8 a AM PM 9 b Hour Min Sec 9 c Month Day Year 9 d ABS REL 9 e C F 10 f TYPE K J or T 10 g OFFSET...

Page 5: ...ures include long battery life and 50 60Hz AC noise rejection This meter is user friendly for field industrial and laboratory applications INITIAL INSPECTION Carefully unpack the unit and accessories...

Page 6: ...cover and make sure to secure the two screws for the water tight feature Note Press the ON OFF key to turn the unit on If the unit is running then you can press the ON OFF key to turn the unit off Th...

Page 7: ...LL mode In the SETTING mode press this key to move back to the previous setting In the RECALL mode press this key to exit the RECALL mode and return to the MEASURE mode CLEAR RIGHT CLEAR In the MEASUR...

Page 8: ...mode When user powers on the model PTM1 the MEASURE mode screen will appear The instrument will return to this mode once you exit from other modes Only in the MEASURE mode the user can press the SAVE...

Page 9: ...e reading it has captured since the last time the instrument has been cleared The instrument will display 2 screens one with time and the other with date The user can press the CLEAR key to clear the...

Page 10: ...o change the flashing value or press the RIGHT key to move to the next date setting After the Year setting press the RIGHT key to move to ABS REL setting Press the LEFT key to move back to the previou...

Page 11: ...s the min max at 03 00 00AM and the last max recorded occurred at 04 00 00AM the next day after more than 1 day the recorded elapsed time will be 01 00 00 e C F Press the UP or DOWN key to change to C...

Page 12: ...e Press the LEFT RECALL key again to exit the RECALL mode and return to the MEASURE mode When the instrument switches to the RECALL mode it will first display the last data saved by the user Three scr...

Page 13: ...ument s range for the thermocouple type sensor used b Probe not properly connected Change sensor with proper temperature range Check probe connection ovEr OVER MAX MAX Unit recorded an over temperatur...

Page 14: ...FS 1 4 F 0 1 FS 0 7 C K 393 F to 2500 F 201 C to 1371 C 1 F 1 C 0 1 FS 1 4 F 0 1 FS 0 7 C T 393 F to 752 F 201 C to 400 C 1 F 1 C 0 1 FS 1 4 F 0 1 FS 0 7 C Cold Junction Compensation Error 0 05 C C 0...

Page 15: ...stomer will be covered by Physitemp Damage due to misuse or accidental breakage is not covered Temperature probes are considered expendable and are not covered by this warranty unless found to be defe...

Reviews: