ERROR DISPLAYS AND TROUBLESHOOTING
LCD
display
Additional
LCD
message
Mode
Possible cause(s)
[Action(s)]
"
ovEr
"
(OVER)
MEASURE
a. Temperature
measured > the
instrument’s range for
the thermocouple type
sensor used.
b. Probe not properly
connected.
[Change sensor with
proper temperature
range.]
[Check probe
connection.]
"
undr
"
(UNDER)
MEASURE
a. Temperature
measured < the
instrument’s range for
the thermocouple type
sensor used.
b. Probe not properly
connected.
[Change sensor with
proper temperature
range.]
[Check probe
connection.]
"
ovEr
"
(OVER)
“
MAX
”
MAX
Unit recorded an over
temperature reading
before.
[Clear the MAX values]
"
undr
"
(UNDER)
“
MIN
”
MIN
Unit recorded an
under temperature
reading before.
[Clear the MIN values]
"
FUL
"
(FULL)
“
SAVE
”
MEASURE
User tries to save data
but the 50 sites are full.
[Go to the RECALL
mode and erase all
data]
“
EPt
”
(EMPTY)
“
RECALL
”
RECALL
User tried to go to the
RECALL mode but no
data has been saved.
[Save at least 1 set of
data]