Basic Settings
8
Basic Settings
The basic settings serve to define different parameters that are rarely modified. You
can get to the basic settings by pressing the
data selector <5>
when switching on
the device with the mains switch.
The basic settings also contain parameters of the
setup menu <10>
.
You can select the following values in the basic settings:
Intensity control, acoustic signal every 0,1 W/cm
2
Dose limit after coupling of ultrasound waves:
YES
d
ose cannot be increased after coupling
NO
dose increment is possible after coupling
Unit of the ultrasonic dose displayed on the
dose indicator <7>:
W or W/cm
2
Link coupling confirmation with timer
YES
Timer runs only with sufficient coupling
NO
Timer runs independently of coupling
Dose threshold, i.e. where instrument switches from continu-
ous to pulsed mode
Volume and pitch of the timer signal
Volume and pitch of the coupling confirmation
Screen contrast
Exit basic settings.