3.2.2 Battery Operation
HIVAMAT
®
200 offers all advantages of battery operation with undiminished possibilities in operation.
How to Start up the Instrument in Battery Mode
(1)
Ensure that the instrument is switched off.
(2)
Turn the
intensity control <6>
to zero.
(3)
Switch on the power with mains switch on the
mains module <1>
. A symbol lights up on the dis-
play of the
timer <5>
. The instrument is in stand-by mode.
(4)
Switch on the instrument by pressing the
massage key <2>
. All displays light up; an automatic self-
test routine is performed by the instrument.
3.3 Function Check
A function check is always necessary when you are unsure about whether the instrument is working prop-
erly.
How to Check the Function of the Instrument
(1)
Ensure that the instrument is switched off.
(2)
Switch on the power with mains switch on the
mains module <1>
. A symbol lights up on the dis-
play of the
timer <5>
. The instrument is in stand-by mode.
(3)
Switch on the instrument by pressing the
massage key <2>
. All displays light up; an automatic self-
test routine is performed by the instrument.
The instrument will automatically carry out a self-test routine during which all functions and output
values are checked. All displays light up; an automatic self-test routine is performed by the instrument
•
If the instrument is properly working, the default values are displayed. You can operate the in-
strument without any further precaution.
•
If an error code is displayed, there is an instrument error. Proceed as described under
on page 12 .
3.4 Cable Check
If you have the impression that a cable is defective during treatment, you can check it using the instru-
ment.
How to Check a Cable
(1)
Switch on the power with mains switch on the
mains module <1>
. A symbol lights up on the dis-
play of the
timer <5>
. The instrument is in stand-by mode.
(2)
Switch on the instrument by pressing the
massage key <2>
. All displays light up; an automatic self-
test routine is performed by the instrument.
(3)
Set the
mode selector <4>
to the middle position.
(4)
Set the
frequency selector <3>
to 50 Hz.
(5)
Set the
intensity control <6>
to the maximum intensity.
(6)
Short-circuit the spiral cable to be tested by plugging one end into the left and the other end into the
right
connector socket <9>
.
PHYSIOMED ELEKTROMEDIZIN AG
3 Instrument Operation
10
HIVAMAT
®
200