Error
message
Problem
Solution
Temperature is less than
32°C or 90°F
Temperature is out of range. Redo test
and make sure temperature is in the range
Temperature higher
than 42.9°C or 109.9°F
Temperature is out of range. Redo test
and make sure temperature is in the range
The thermometer is
moved when measuring
Redo the test. Do not move when
measuring.
The time of predictive
measurement is excess
10 seconds.
The Ambient temperature may be too low,
please measure at higher temperature.
Person may just drink a cold beverage;
please repeat measurement in
15 minutes.
The probe tip may not have cooled down,
please wait 30 seconds between each
measurement.
The thermometer is moved when
measuring. Do not move, and then redo
the test.
Other Err
message
The thermometer is not
functioning properly.
Unload the battery, wait for 1 minute
and repower it. If the message reappears,
contact AMG Medical customer service.
Direct Current
Batch Code
Type BF Applied Part
Consult Accompanying
Documents
Type:
Digital Thermometer (Predictive)
Measure Range:
32.0°C - 42.9°C (90.0°F - 109.2°F)
Accuracy:
±0.1°C(0.2°F) between 35.5°C~42.0°C
(95.9°F~ 107.6°F) at 18°C~28°C(64.4°F~82.4°F)
ambient operating range
±0.2°C (0.4°F) for other measuring and ambient
operating range
Operating mode
Predictive mode: oral, rectal, underarm
Direct mode (real temperature): bath
Display:
Liquid crystal display, 3 1/2 digits
Memory:
For storing the last measured value
Battery:
One 3.0 V DC. button battery (size CR2032)
Battery life:
Approximately 200 hours
Dimensions:
13.5 cm×3.4 cm×1.7 cm (L×W×H)
Weight:
Approx. 25 grams including battery
Ambient operating range:
Temperature: 5°C~40°C (41°F~104°F)
Relative humidity: 15% ~ 95%RH
Storage and transportation
condition:
Temperature: -20°C ~ 55°C(-4°F~131°F)
Relative humidity: 15% ~ 95% RH
Ingress Protection Rating:
IP27
Classification:
Classification: Type BF