10
TrueCPR Device Instructions for Use
Using the TrueCPR Device
Shut-down Screen
Event Review Screens
Two event review screens display data for the most recent device use. To display the event review
screens, start with the device turned off. Press the
(AIRWAY)
and
(POWER)
buttons simultaneously to
turn the device on in Event Review mode. To switch between the screens, press the
(MUTE)
button.
Progress bar appears during shut-down.
Note:
Shut-down may take several minutes. Do not
remove batteries while shut-down is in progress.
Percent of compressions with depth > 5 cm (2 in).
Percent of compressions with proper recoil.
Average compression rate (compressions per minute).
Total event duration.
Percent of total event with compressions.