Battery Status: To check battery status for non-connected devices, refer to
•
90 days from Battery Expiration: Device readiness status changes to
Needs Attention - Battery expiring
which will trigger an email notification being sent to customer with email title:
Needs Attention - Battery
expiring
. The email will be resent monthly if device status does not change.
•
After Battery Expiration: Device readiness status changes to
Not Ready (Battery expired)
and an email
notification is sent. Email notification will be resent weekly if device status does not change (Monday at
12:01am local time).
Performance Inspection Procedure (PIP)
Overview
: The Performance Inspection Procedure (PIP) is a set of manual and computer assist test procedures. The
PIP is used for an operational, closed-case evaluation of the LIFEPAK CR2 defibrillator to determine if the defibrillator
is operating within the required specifications. The PIP will be performed in a regularly scheduled preventive
maintenance routine.
NOTE
:
1. Always use the latest version of the PIP document, PN 3323342, to perform PIP.
2. PIP document, PN 3323342, can be downloaded from SmartSolve.
3. Fillable PDF format PIP checklist, 3323342_F_PIP_Checklist_LPCR2.pdf, can be downloaded from
SmartSolve.
Replacement Procedure
Supply and Replacement Parts
The following table is the list of supplies and accessories that are available for replacement.