LIFEPAK CR Plus/EXPRESS Defibrillator
Preventive Maintenance
Replacing the CHARGE-PAK Battery Charger (continued)
3. Discharge the used battery charger using the discharge tool that is supplied
with the replacement kit
4. Insert the new battery charger into the defibrillator until you hear it “click” into
position. If the battery charger is not completely inserted, it will spring back
out.
5. Confirm that the
BATTERY
indicator disappears and that the
OK
indicator
appears on the Readiness Display before returning the defibrillator to
service. (The internal battery must charge for a period of time before the
OK
indicator appears.)