LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Troubleshooting Chart
(continued)
Area
Observed Symptom
Suggested Corrective Action
Audio
Inaudible or garbled audio
Perform the voice tone test.
Check the speaker connection.
Check or
replace the W02 Speaker Assembly
.
Power source
management
No backup battery operation
Therapy - delivered
energy
Unable to complete auto test
or user test
Rerun the test with proper test load shorting.
Check continuity of the test plug or shorting bar.
Check continuity of the QUIK-COMBO cable.
Check or
replace the W01 Therapy Connector
.
Delivered energy out of
tolerance
Perform defibrillator calibration.
Patient impedance
channel broken
Abnormal energy delivery
replace the A14 Inductive Resistor
5 ohms).
Check or
.
.
(Continued on next page)