Setup Options
8-6
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
To activate, select
ENERGY PROTOCOL
in the Paddle Default menu. Auto energy sequences
are disabled if you press the
ENERGY SELECT
control or change to or from AED mode during
use.
Table 8-4
Manual Mode Energy Protocol Setup Menu
Menu Item
Help Message
Options
PRESET
PROTOCOL
Select preset energy
protocol
FULL RANGE
,
PEDIATRIC
.
ENERGY 1
Select energy level for
shock 1
Full range:
100, 125, 150, 175
,
200
,
225,
250, 275, 300, 325, 360
.
Pediatric:
2, 3, 4
,
5
,
6, 7, 8, 9, 10, 15, 20, 30,
50, 70, 100
.
ENERGY 2*
Select energy level for
shock 2
Full range:
100, 125, 150, 175, 200, 225,
250, 275
,
300
,
325, 360
.
Pediatric:
2
,
3
,
4
,
5
,
6
,
7
,
8
,
9
,
10
,
15
,
20
,
30
,
50
,
70
,
100
.
ENERGY 3*
Select energy level for
shock 3
Full range:
100, 125, 150, 175, 200, 225,
250, 275, 300, 325
,
360
.
Pediatric:
2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30,
50, 70, 100, 125, 150
.
*
ENERGY 2
cannot be less than
ENERGY 1
.
ENERGY 3
cannot be less than
ENERGY 2
.
Summary of Contents for LIFEPAK 20
Page 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Page 3: ......
Page 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Page 15: ......
Page 49: ......
Page 111: ......
Page 155: ......
Page 171: ......
Page 181: ......
Page 183: ......
Page 189: ......
Page 191: ......
Page 195: ......
Page 199: ......
Page 201: ......
Page 205: ......
Page 209: ......
Page 211: ......
Page 213: ......
Page 215: ......
Page 221: ......
Page 226: ......