LIFEPAK 15 Monitor/Defibrillator
Service Manual
Modes of Operation
Service Mode
95
4
Entering Service Mode
To enter the Service mode:
1.
Enter the Setup Mode (see
)
2.
Rotate the SPEED DIAL to select SERVICE in the Setup
menu, and then press the SPEED DIAL. The Service mode
passcode prompt appears. The factory default passcode is
0000; the reserved technician passcode is 5433.
3.
To enter the passcode, rotate the SPEED DIAL to select a
digit, and then press the SPEED DIAL to continue. After the
last digit is entered, the Service menu appears.
4.
Rotate the SPEED DIAL to select a service option, and then
press the SPEED DIAL to display the option overlay.
Calibration
— See Test and Calibration Procedures.
Tests
— See Performance Inspection Procedures.
Status
— See
Set Passcode
— Set a Service mode access passcode.
Maint Prompt
Preventive Maintenance (p. 145)
.
Setup
— Return to the Setup Home Screen.