Batteries
34
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
To install a battery:
1.
Confirm that the battery is fully charged, unless the battery will be charged in the
monitor/defibrillator using the power adapter.
2.
Inspect battery pins and contacts in the battery wells for signs of damage.
3.
Align battery so battery clip is over the pins in the battery well.
4.
Insert the end of the battery that is opposite the battery clip into the battery well.
5.
Firmly press the clip end of the battery into the battery well until it clicks into place.
6.
Repeat Step 1 through Step 5 to insert second battery.
To remove a battery, press the battery clip in and tilt the battery out of the battery well.
Possible Loss of Power During Patient Care
Battery pins in the defibrillator may be damaged if
batteries are dropped or forced into battery wells.
Inspect pins routinely for signs of damage. Keep
batteries installed at all times except when the device is
removed from service for storage.
For information about battery maintenance, see Battery Maintenance (on page 209).
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......