System Description
3.8
DIP Switch Settings
Switch 1 to 4:
Address / device number (16
possible combinations)
Switch 5 and 6:
Baud rate
Switch 7
Limit switch mode
Switch 8
Firmware update mode
Fig. 8: slider up ON
slider down OFF
Factory settings are shown in bold.
Dev.
No.
Switch
Value
SW1 SW2 SW3 SW4
1 0
ON
ON
ON
ON
2 1
ON
ON
ON
OFF
3 2
ON
ON
OFF
ON
4 3
ON
ON
OFF
OFF
5 4
ON
OFF
ON
ON
6 5
ON
OFF
ON
OFF
7 6
ON
OFF
OFF
ON
8 7
ON
OFF
OFF
OFF
9 8
OFF
ON
ON
ON
10 9
OFF
ON
ON
OFF
11 10
OFF
ON
OFF
ON
12 11
OFF
ON
OFF
OFF
13 12
OFF
OFF
ON
ON
14 13
OFF
OFF
ON
OFF
15 14
OFF
OFF
OFF
ON
16 15
OFF
OFF
OFF
OFF
Baud rate*
SW5
SW6
9600
ON
ON
19200 OFF
ON
*Other settings are fixed at 8 data and 1
stop bit, no parity; Internal buffers are
used so there is no handshake required;
MMCRun
supports only 9600 baud
HW Limit Switch Mode* SW7
Limits active-low
ON
Limits active-high
OFF
* Hardware limit switch mode must agree
with software setting
Firmware Update Mode
SW8
Update
ON
Normal operation
OFF
NOTE
In the native command set the controller’s “address” is the same as the
“switch value”.
In the GCS command set, the controller’s “address” is the same as the
“device number”.
www.pi.ws
C-863 MS173E Release 1.2.7
Page 23