There are following limitations in SYNC OUT TIMES function
Frame Rate
Restriction
~
60,000fps No
Limit
60,001fps
~
90,000fps
×
30 Nonusable
90,001fps
~
500,000fps
×
20
、
30 Nonusable
500,001fps
~
700,000fps
×
8
、
10
、
20
、
30 Nonusable
700,001fps
~
775,000fps
×
6
、
8
、
10
、
20
、
30 Nonusable
3.18. Event Marker Function
With the system, it is possible to input an external signal during recording, at the instant the frame
number is stored, and during playback you can immediately access, or jump to, the stored frame
numbers (event markers). This a separate feature from the trigger point, by marking interesting
points during recording, these points can be easily called up for review during playback. The event
marker can store ten positions within a sequence.
The frame number recording occurs on the pulse's edge, and the next frame after the pulse's edge is
input is stored as the event marker. Event marker settings can be made with the "LCD remote