FASTCAM SA-X2
28
1.3 Input / Output Signal T
ype
s
1.3 Input / Output Signal Types
With the system, many signals can be input and output through the I/O cable. Signals that can be
input and output from the I/O cable are listed below.
1.3.1
TRIG TTL IN Connector
The system recognizes an external TTL signal as a trigger during the READY or ENDLESS recording
state. Starting and stopping recording (in the selected recording mode) is controlled with this
signal.
Input voltage is 0V to +12V (H level +2.5V to +12V), positive or negative polarity, pulse width is
200 ns or greater.
1.3.2
TRIG SW IN Connector
This trigger is input during the READY or ENDLESS recording state by contact between the BNC
connector's shield and a center pin (switch closure). The center pin normally has voltage flowing
through it. Use caution to avoiding contact with other pins.
A signal other than the specified signal must not be input to the various connectors.
Use extreme caution as there is a risk of damage to both, the input device and the output
device.
Connector Name
(Input System)
Menu
Signal
TRIG TTL IN
TRIG POS
FET Input 0V - +12V (H level +2.5V to +12V), Positive Polarity
TRIG NEG
FET Input 0V - +12V (H level +2.5V to +12V), Negative Polarity
Connector Name
(Input System)
Menu
Signal
TRIG SW IN
None
Contact signal
Summary of Contents for FASTCAM SA-X2
Page 1: ...FASTCAM SA X2 Hardware Manual Rev 4 00E...
Page 41: ...Chapter 2 Recording This chapter explains operations related to recording...
Page 46: ...Chapter 3 Product Specifications This chapter explains the system s specifications...
Page 67: ...Chapter 4 Warranty This chapter explains about the warranty...
Page 71: ...A Appendix...