Chapter. 3
Functions
3.16.4. Synchronizing Multiple FASTCAM Mini WX50/100 Systems
(Multiple Unit Synchronized Recording)
The system can perform synchronized recording by synchronizing multiple units using external.
Synchronization input/output
Synchronized recording settings using the system are made with PFV. The conceptual settings
when performing synchronized recording using two systems are explained here. First, decide which
camera to make the master camera (outputs the synchronization signal) and the slave camera
(receives the synchronization signal) from the two systems to use for synchronized recording.
Setting the Master Camera (Outputs Synchronization)
Set the signal output for the master camera which will output the synchronization signal. Synchronization
signal settings are made with PFV.
1.
Verify that the camera mode is in LIVE mode (the image displayed is passed through from the
camera). If the system is in a mode other than LIVE mode, check "Live" on the camera
control panel.
2.
Select I/O on the left tree from "Camera Option" on the camera control panel.
3.
Ensure that “OUTPUT1” is set “SYNC POS”.
4.
Ensure that “OUTPUT2” is set “TRIG POS”.
OUTPUT 1
(MASTER)
(SLAVE)
CAMERA No.1
CAMERA No.2
(BNC Cable)
OUTPUT 2
INPUT 2
INPUT 1
50
Summary of Contents for FASTCAM Mini WX100
Page 1: ...F FA AS ST TC CA AM M M Mi in ni i W WX X5 50 0 1 10 00 0...
Page 12: ......
Page 26: ...Chapter 1 Overview 14...
Page 40: ...Chapter 2 Setup 28...
Page 73: ...61 FASTCAM Mini WX50 100 Hardware Manual...
Page 76: ...Chapter 5 Product Specifications 5 1 Specifications 5 2 Dimensions 64...
Page 80: ...Chapter 5 Product Specifications 68...
Page 94: ...Chapter 5 Product Specifications 5 2 2 AC Power Supply Unit mm 82...
Page 95: ...83...
Page 96: ...Chapter 6 Warranty Chapter 6 Warranty 6 1 About the Warranty 84...
Page 98: ...Chapter 7 Contacting Photron Chapter 7 Contacting Photron 7 1 Contact Information 86...