34
SenSys User Manual
CCD Specifications
KAF 0400
H x V: 768 x 512
Pixel Size: 9µm x 9µm
Typical QE Curve (not installed in camera)
QE response in the 200-400 nm range can be enhanced with Photometrics’ proprietary Metachrome II coating.
Detection Mode
Gain
Setting
Relative
Gain
Typical
Noise
System Gain
Typical Maximum ADC
Signal
High Signal to
Noise
1
≈
1
/
2
x
26e
-
≈
40e
-
/ADU
160Ke
-
(binned full well*)
High Dynamic
Range
2
≈
1x
19e
-
≈
20e
-
/ADU
80Ke
-
(single pixel full
well)
High Sensitivity
3
≈
4x
13e
-
≈
5e
-
/ADU
20Ke
-
(high sensitivity)
* Binning must be a minimum of 2 pixels in the parallel and serial direction to reach the maximum ADC signal.
400
500
600
700
800
900
1000
Wavelength (nm)
0
10
20
30
40
50
Quantum Efficiency %