HB595/A
TANNING MEASUREMENT
- The measuring data indicated below where established with help of an UVX 36 meter and may only be checked with this
equipment, since the values measured are relative values.
- The measurements must be taken after approx. 5 minutes in the centre of the tanning field and at the distance indicated.
- The following aspects must be checked or the following preliminary measurements must be taken:
• Mains voltage: 230 V, 50 Hz
• Room temperature 22 - 28 °C
• Check the fans for failure-free operation as optimal tanning is only guaranteed at the correct operating temperatures
• Clean the lamps, the acrylic sheets and the glass pane.
- With regard to the minimum values indicated below, you should bear in mind that this is not an absolute value at which the
lamps must be replaced.
- Depending on the skin type, the tanning session may also be extended so that the lamps can be used a little longer.
UVX 36 meter
distance in cm
output in mW/cm²
NEW
(approx.)
Output in mW/cm²
MIN
(approx.)
Couch
2.5 cm from acrylic sheet couch
9.5
6.6
Canopy
30 cm from acrylic sheet couch
10.0
7.0
Canopy beneath the
facial tanner
2.5 cm from acrylic sheet couch, and
on 14 cm from edge of the end cover
5.2
3.6
Note:
Switch SK1 of facial tanner in position ’HIGH’.
4-11