Alarms, messages, and troubleshooting
9-17
Vent Inoperative
1005
Internal Tempera-
ture High
Technical failure of the
motor PCBA. The venti-
lator is in the ventilator
inoperative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(7)
No
No
No
Vent Inoperative
1006
Data Acquisition
PCBA ADC Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(8)
No
No
No
Vent Inoperative
1007
Machine and
Proximal Pres-
sure Sensors
Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(9)
No
No
No
Vent Inoperative
1008
Machine and
Proximal Pres-
sure Sensors
Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(10)
No
No
No
Vent Inoperative
1009
Pressure Regula-
tion High
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(11)
No
No
No
Vent Inoperative
100A
Data Acquisition
PCBA ADC Refer-
ence Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(12)
No
No
No
Vent Inoperative
100B
Watchdog Test
Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(13)
No
No
No
Table 9-5: Vent Inoperative alarm messages: summary and troubleshooting (continued)
Message
Description
Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...