Note: Dirt, grease or particles left on the skin may
lodge themselves in the hydrogel of the electrodes. This
has an adverse effect on the adhesive power of the
electrodes.
3
Take the electrodes from the plastic bag and
check them for damage and signs of wear.
Caution: Do not use electrodes whose use-by
date has expired. The use-by date is shown on the
resealable plastic bag.
Note: Replace the electrodes when they are damaged
or dirty, when they have lost their adhesive power or
when stimulation becomes uncomfortable, i.e. when you
experience an unpleasant stinging or biting sensation.
4
Pull the electrode off the liner and place it
on the skin at the place you want to treat.
See section ‘Electrode placement’ and the
placement guide.
5
Place the second electrode on the body part
you want to treat.
Note: Always place the electrodes on the skin before
you connect the device.
Note: Do not place the electrodes on top of each other
or so close to each other that they touch each other.