Problem
Possible cause
Solution
The device makes a ticking
sound after disconnecting
the device from the wall
socket.
The timer is not in the off
position.
Turn the timer to the off position.
The device switches off
before the preset
treatment time has
elapsed and sometimes it
automatically switches on
again.
The cooling vents are blocked
and the device becomes too
hot.
Make sure the cooling vents are not
covered or blocked.
The fan is defective and the
device becomes too hot.
Contact the Philips Consumer Care
Center.
Specifications
Model
PR3140
Rated voltage (V) and rated frequency (Hz)
220-240 V ~50-60 Hz
Rated consumed power/output power
650 W
Medical device classification
IIa
Insulation class
Class II
Ingress of water
IP21
IR type
A/B/C
Irradiance at 30 cm from the filter
approx. 1000 W/m
2
Service life
5 years
Mode of operation
Continuous
Net weight
Approximately 7 kg
External dimensions
263x394x833 mm (LxWxH)
Infrared light
Tested according to
IEC 60601-2-57:2011
Wavelength
700-3000 nm
Max output
2.8 kW/m
2
Irradiance
2.8 kW/m
2
Radiance
31
kW/sr/m
2
Maximum variation of the output from the mean
value across the treatment area
480
W/m
2
15
English