2 .1. Installation
2.1.1. POSITION
The instrument may be used in any position except on its rear end. Make sure that the i nstrument is not
exposed to excessive heat.
2. 1.2. CONNECTION TO THE MAINS
The instrument must have a.c. supply only. On delivery, it is set for a mains voltage of 220 V. Setting for a
d ifferent mains voltage is done as follows:
- Remove the left screw of the handle bar and take off the l eft side-pl ate.
- Change the primary connections of the mains transformer accord ing to Fig. 3, a diagram of which is in the
inside of side-plate.
-
Change the mains voltage ind ication-plate at the rear.
- Refit the hand l e bar.
- Connect up the i nstrument.
2.1.3. EARTHING
The instrument must be earthed i n accordance with the local safety regu lations. The ma ins cab l e suppl ic3d is
prov ided w ith an earth core which is connected to the earth contacts of the mains plug. If the instrument is
con nected to a mains socket witf-J earth contacts, the cabinet is automatically earthed.
The c ircuit earth of the i nstrument and the chassis-connection of the BI'JC-connectors are at chassis potential.
Earthing the instrument via the chassis-connection of the BNC-connectors is not permitted.
5
1
V
Fig.
3.
Changing the primary connections of the mains transformer
Summary of Contents for PM5324
Page 1: ...HF Generator 100kHz 110MHz PM5324 9452 053 240 1 9499 450 05511 730801 1 01 ...
Page 15: ...14 Fig 4 Front view Fig 5 Rear view ...
Page 43: ...Fig 33 Front view with item numbers Fig 34 Bottom view ...
Page 44: ...44 Fig 35 Top view Fig 36 Left side view ...
Page 45: ...45 0 0 0 0 1 ji I I MAB225 Fig 37 Scale dri ve n ...