Telemetry Functions at the Information Center
9-6
Telemetry Functions at the Information Center & TeleMon
Patient
Configurable
Settings in
Telemetry
Setup
The Telemetry Setup window enables you to configure the transceiver for
patient-specific settings. All patient-specific settings will be reset to the unit
defaults upon patient discharge. To access the window, from the Patient
Window click
Telemetry Setup
.
The following settings can be adjusted in this window.
Patient-Configurable Settings in Telemetry Setup
Control
Function
Setting Choices
Factory
Default
Telemetry
Button
Determine the Information Center
response when Telemetry Button
is pressed.
Nurse Call -
generate nurse
call alarm
Record
- generate a
recording strip
Nurse Call
and
Record
-
generate nurse call alarm and
recording strip
None
Nurse Call
Telemetry
Device:
Volume
Set the volume level for all
adjustable sounds on the
transceiver.
1
(low),
2
,
3
,
4
,
5
(high)
3
Telemetry
Device:
Mute
Enable/disable Spot Check and
pulse detection tones on the
transceiver.
enable
(mute)
disable
enable
Fixed Pacer
Amplitude
Sets the appearance of the pacer
spikes to a fixed size as they
appear in the patient window.
enable
disable
disable
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...