When
Connected
to TeleMon
When the transceiver is connected to TeleMon:
•
The SpO
2
measurement mode is
always
Continuous.
•
You can change the mode. Changes to the mode take effect when the
transceiver is disconnected from TeleMon. The following settings will be
used:
Turning SpO
2
Monitoring Off
To turn SpO
2
monitoring off, disconnect the sensor cable from the transceiver.
SpO
2
enters a power-down mode after the cable is disconnected from the
transceiver, thereby conserving battery life.
It is important to disconnect the sensor from the transceiver in order to conserve
battery life. Unplugging the sensor from an adapter cable that is connected to
the transceiver does NOT provide SpO
2
power-down mode.
SpO
2
should also be turned off at the Information Center. If the transceiver is
configured for Continuous SpO
2
measurement and the sensor is removed
without turning SpO
2
off, a SpO2T NO SENSOR technical alarm will result.
Silencing this technical alarm will turn the SpO
2
measurement off.
Mode Set at TeleMon
Mode when Transceiver
is Disconnected
Continuous
Continuous
5-minute
Continuous
1-minute
Continuous
Manual
Spot Check
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...