Measuring SpO
2
SpO
2
Monitoring
8-17
Spot Check
Measurement
When the transceiver is configured for Spot Check measurement, use the
following instructions to take an individual, manual SpO
2
reading from the
transceiver.
Note—
You can also initiate a Spot Check measurement in the Patient Window
at the Information Center by selecting the
Spot Check SpO
2
icon (see
“Patient Configurable Settings in Telemetry Setup” on page 9-6).
Note—
Spot Check measurements cannot be made when the transceiver is
connected to TeleMon because only Continuous measurement is available with
this device.
Step
Action
1
Attach the sensor to the patient.
2
Connect the SpO
2
cable to IntelliVue TRx
+
, and check that:
•
The SpO
2
sensor light turns on.
•
A low-pitch tone detecting each pulse is audible (unless
sounds are muted).
3
After approximately 30 seconds, if the measurement was successful,
you’ll hear a single beep from the transceiver. The value, with the
measurement time, will be displayed at the Information Center. The
sensor light will extinguish.
If the measurement was unsuccessful, you’ll hear a double beep.
Remove the sensor cable and reinsert it to retake the measurement.
Note—
The SpO
2
value and time stamp remain on the Information
Center for 1 hour or until another measurement is taken, with one
exception: If the batteries are removed from the transceiver, the Spot
Check measurement will be erased from the display; however, the
SpO
2
measurements will be available in Trend Review.
4
To repeat a Spot Check
measurement at the bedside, disconnect
then reconnect the SpO
2
cable to the transceiver.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...