Opening the Sealed Case
4
-
28
Removal and Replacement
Secondary Method
a.
Connect clip leads to each end of one of the Disarm Resistors.
b.
Connect the clip leads directly to the spade terminals on the Power
PCA where the Defibrillator Capacitor is connected. Maintain this
connection for at least 5 seconds.
c.
Repeat steps a and b using the other Disarm Resistor.
The defibrilla-
tor capacitor is now discharged, even if one of the resistors is dam-
aged.
Figure 4-16 Discharging the Defibrillator Capacitor
High Voltage
Discharge Tool
Touch here first
Then touch here
Capacitor & cover
Inductor
Disarm resistors