41
handling, the pad is designed with a non-gel
area around the connector cable.)
device operation
Defibrillation may not be provided if a user
delays pressing the Shock button. The
HeartStart will only deliver a shock if the
flashing orange Shock button is pressed
when the instruction is given. If the Shock
button is not pressed within 30 seconds
after the instruction, the HeartStart will
disarm itself, and (for the first CPR interval)
give a reminder to make sure emergency
medical services have been called. The
HeartStart will then begin a CPR interval.
This is designed to minimize interruption of
CPR and help ensure ongoing patient
support.
If for any reason you want to turn off the
defibrillator during a use, you can press the
On/Off button – holding it down for at least
one second
-
to return the device to
standby mode.
HS1_Home_OM_English.book Page 41 Thursday, June 13, 2019 1:18 PM