output in
output in
UVX 36 meter
distance in cm
mW/cm2
mW/cm2
NEW
MIN
(approx)
(approx)
Couch
2.5 cm from acrylic sheet couch
10.0
6.3
Canopy
30 cm from acrylic sheet couch
10.0
6.3
Canopy beneath the facial tanner
30 cm from acrylic sheet couch
6.1
3.9
Note: Switch SK1/SK2 of HP lamp (facial tanner) in position 2.
IRRADIANCE
The measuring data indicated below were
established with help of an UVX 36 meter and
may only be checked with this equipment, since
the values measured are relative values.
The measurements must be taken after approx.
5 minutes in the centre of the irradiation field and
at the distance indicated.
The following aspects must be checked or the
following preliminary measurements must be
taken:
• Mains voltage: 3 x 230V
• Room temperature approx.: 22-28°C
• Check the fans for failure-free operation as
optimal irradiation is only guaranteed at the
correct operating temperatures.
• Clean the reflectors, the lamps, the acrylic
sheets and the glass pane.
With regard to the minimum irradiation indicated
below, you should bear in mind that this is not an
absolute value at which the lamps must be
replaced.
Depending on the skin type, the tanning session
may also be extended so that the lamps can be
used a little longer.