FC6844
DISASSEMBLY- AND RE-ASSEMBLY ADVISE
Motor unit:
- To open the motor unit,
remove the inlay by pressing
the inlay on the cord winder
push-button and remove 2
screws (
A
). (The inlay is
mounted on the cover by
means of a snap
construction).
- Use a long Philips screwdriver
9965 000 08332
to remove
the 6 screws (
B
) of the lower housing / top housing.
(Holes in lower housing are 13 cm deep).
Accessories:
- See drawing for replacing of the shampoo pressure hose, for
example the pistol handgrip.
1
2
2
1
3
4
5
REPAIR INSTRUCTION
- The type no. plate is placed on the outside of the lower
housing (
Item 38
).
- For safety reasons, make sure that the cap clamps do not
come above the panel surface (
Item 31
) after replacing of
the switch (
Item 30
).
If the lower housing (
Item 38
) must be separated from the
top housing (
Item 15
)
always
replace waterproofing foam
(
Item 22
). Press hard on the housing during assembly of
this unit to prevent the foam from moving during assembly.
- The cord winder lever
(
Item 14
) must be assembled
through the top cover at an
angle (as shown in the figure)
to operate the cord winder.
- Depending on the national version, the cleaner has a 6.3
or 9.3 m cord. The spring unit (
Item 101
) is standard
provided with a 9.3 m spring.
- To demount the cord winder use TORX screw driver no.
T15, which is one of a set consisting of T20, T15, T10, T9,
T8 and T6 4822 395 50145.
- To check the shampoo circuit, it is neccessary to use the
male plug (
Item 61
) to open the valve in the female plug
(
Item 32
).
If the real half spring unit (
Item 101
) is replaced, shaft
E
must be lubricated with
silicone grease
. Do not apply any
silicone grease to or spill any grease on terminals (
Item 103
)
or conductors (
Item 102
).
30˚
±
2
1
A
A
2-6