To access provider mode:
1. From the main menu, highlight
Setup
.
2. Press and hold ramp button and control wheel
simultaneously for 4-5 seconds until beep.
3. Provider
Setup
screen then appears.
You will remain in provider mode until you exit the Setup
screen using the
Back
selection, or until a screen time-out
occurs after one minute.
To program device:
1. Turn control wheel to specific setting and press.
2. Turn control wheel to desired option and press control
wheel to select.
3. Turn control wheel to next setting and repeat.
4. To exit, turn control wheel to
Back
and press.
Provider setup menu
Therapy settings
Mode
CPAP, S, S/T, T, PC
CPAP
4-20 cmH
2
O
AVAPS
On, Off
Vte
200-1500 ml
IPAP max
4-25 cmH
2
O
IPAP min
4-25 cmH
2
O
IPAP
4-25 cmH
2
O
EPAP
4 to IPAP
BPM
0-30
Ti
0.5-3.0 sec
Comfort settings
Flex control
On, Off
Flex setting
Off, 1, 2, 3
Rise time control
On, Off
Rise time
1-3 (1=200 ms; 2=300 ms; 3=400 ms)
Ramp time
0-45 min
Ramp start
4 to EPAP
Tubing type control*
On, Off
Tubing type*
15, 22
*for firmware version 2.11 or higher
Set up device