22
English
Single electric breast pump
Double electric breast pump
Storage and transportation
conditions
Temperature: -25 °C to 70 °C (-13 °F to 158 °F)
Relative humidity: up to 90 % (non-condensing)
Operating and storage pressure
700-1060 hPa (<3000 m altitude)
Net weight
230 g
310 g
External dimensions
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Device classification
Applicable classifications:
Degree of protection against electric shock:
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical system
according to IEC 60601-1.
Materials
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
Service life
500 hours
Mode of operation
Continuous operation
Ingress protection
IP22 (motor unit), IP20 (adapter)
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic compatibility
(EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables. Only use this Philips
Avent single/double electric breast pump Plus in combination with the adapter supplied.
Emissions
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The device does not use RF energy for its intended
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power supply
network which supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
CompliesClass A Category
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer‘s declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Plus is intended for use in the electromagnetic environment listed
below. The user of the device should ensure that it is used in such an environment to prevent improper operation.
Summary of Contents for Avent SCF391