11 Upgrades
215
Installing Upgrade Options
This section covers how to install the options.
Option B71
To install the noninvasive blood pressure upgrade:
1
Remove the top cover assembly as described in “Removing the Top Cover Assembly” on
page 141.
2
Remove the front end assembly (see “Removing the Front End Assembly” on page 200).
3
Remove the connector block frame from the front end assembly (see “Disassembling the Front
End Assembly” on page 202). Leave the fetal sensor socket connector block in place.
4
Take the new noninvasive blood pressure socket. Locate the metal pin and the plastic stud on the
underneath of the noninvasive blood pressure socket in the corresponding holes in the bus master
board.
J70
System Interface, 1 x RS232
port and 1 x LAN port.
RS232/LAN I/O card
assembly
M2703-67501
"Plug & Play" interface
1
SJ3
Rel. J.3 Software Upgrade
Fetal Monitoring
Documentation DVD-
ROM
453564491101
-
1
Instructions for Use
453564490751
Appropriate localized version
1
FM20/30 Upgrade Options
Option
Number
Option Adds
Parts included with Option
Description
Part Number
Contents/Comments
Qty
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...