11 Monitoring Twin FHRs
115
Monitoring Twins Internally
FM30/50
Monitor one twin using the procedures described in the Chapter “Monitoring FHR and FMP Using
Ultrasound” on page 95. Monitor the second twin using the procedures described in the Chapter
“Monitoring FHR Using DECG” on page 137.
Example of a screen showing twin monitoring using a combination of US and DECG (The fetal heart
rate monitored via DECG is labeled "DFHR1"/"DFHR2"/"DFHR3" on the screen):
Cross-Channel Verification
If the monitored heart rates (from a fetal or maternal source) coincide at any time (that is, if the same
heart rate is being monitored by more than one transducer), this is detected via the monitor's cross-
channel verification feature, and
appears on the screen and is repeatedly printed on the trace paper
after about 30 seconds. If the suspended coincidence persists for more than one minute without
interruption, then the INOP
COINCIDENCE
will additionally appear on the screen. If you are
monitoring externally, check the trace and reposition one of the transducers, if necessary, to detect the
second FHR correctly.
FM30/50
If you are monitoring internally, check the trace and, if necessary, reposition the ultrasound transducer
to detect the second FHR correctly.
1
DFHR 1
2
Toco parameter
3
FHR 2
Summary of Contents for Avalon FM20
Page 2: ......
Page 8: ...8 ...
Page 56: ...3 Basic Operation 56 ...
Page 72: ...5 Alarms 72 ...
Page 82: ...6 Patient Alarms and INOPs 82 ...
Page 86: ...7 Admitting and Discharging 86 ...
Page 90: ...8 Non Stress Test Timer 90 ...
Page 120: ...11 Monitoring Twin FHRs 120 ...
Page 126: ...12 Monitoring Triple FHRs 126 ...
Page 136: ...15 Monitoring Uterine Activity Internally 136 ...
Page 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Page 154: ...18 Monitoring SpO2 154 ...
Page 164: ...20 Printing the ECG Waveform 164 ...
Page 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Page 170: ...22 Recovering Data 170 ...
Page 174: ...23 Care and Cleaning 174 ...
Page 188: ...25 Accessories and Supplies 188 ...
Page 216: ...26 Specifications and Standards Compliance 216 ...
Page 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...