PHILIPS Respironics Actiwatch 2 Technician Manual Download Page 10

10

Recommended Separation Distance Between Portable and Mobile RF Communication Equipment and the Device.

The Device is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The user 
of the Device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile 
RF communication equipment (transmitters) and the Device as recommended below, according to the maximum output 
power of the communications equipment.

Rated Maximum 
Output Power of 
Transmitter (Watts)

Separation Distance According to Frequency of Transmitter (meters)

150 kHz to 80 MHz

d = 1.2√P

80 MHz to 800 MHz

d = 1.2√P

800 MHz to 2.5 GHz

d = 2.3√P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) 
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power of 
the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 

from structures, objects and people.

Summary of Contents for Actiwatch 2

Page 1: ...Actiwatch Technician Guide...

Page 2: ...wearer shows signs of skin reddening or inflammation If the device becomes damaged discontinue use and return it for replacement or repair Do not connect the device to a computer while worn or handle...

Page 3: ...uide 10 pack Technician Clinician Guides Actiwatch 2 Communication Dock USB Wall Plug with Cable Actiware Software Replacement Watchbands andTool Wearer Guide 10 pack Technician Clinician Guides Symbo...

Page 4: ...e Database New 3 For each new patient create a new patient record with a unique Subject ID in the database by selecting File New Subject 4 Connect your device to your computer with the provided USB ca...

Page 5: ...Note Only Actiwatch 2 Spectrum PRO and Spectrum Plus devices are rechargeable Return Actiwatch Spectrum devices to Philips Respironics for battery replacement Actiwatch Spectrum PRO and Plus Connect y...

Page 6: ...watch is five years WEEE Recycling Directives Separate collection for electrical and electronic equipment per EC Directive 2002 96 EC Dispose of this device in accordance with local regulations Contac...

Page 7: ...ts for Safety IEC 60601 1 2 2007 Medical Electrical Equipment Part 1 2 General Requirements for Safety Collateral Standard Electromechanical Compatibility Requirements andTests IEC 60601 1 11 2010 Med...

Page 8: ...magnetic environment specified below The user of the Device should assure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidanc...

Page 9: ...s determined by an electro magnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbo...

Page 10: ...nsmitter Watts Separation Distance According to Frequency of Transmitter meters 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1...

Page 11: ...s or workmanship This warranty does not cover damage that may occur in shipment This warranty does not apply to any Product or individual parts that have been repaired or altered by anyone other than...

Page 12: ...1104768 1104715 R01 JDH 04 24 2013...

Reviews: