background image

LD60A 

8-1 

SECTION 8:  PARTS LIST 

 

 

 

 

 

 

ITEM 

P/N 

DESCRIPTION 

QTY 

 

CB1 

1601469 

CIRCUIT BREAKER 

CON1 

POWER ENTRY MODULE 

CON2 

1159900 

AC OUTLET 

CORD 

1077183 

POWER CORD, 8' 

DAQ1 

WIRELESS MULTIFUNCTION DAQ DEVICE 

F1 

1603632 

FUSE 

F2A, B 

1603704 

FUSE 

KNOB 

1355310 

KNOB 

LED1 

LED, RED 

LED2 

LED, YELLOW 

LED3-4 

LED, GREEN 

METER 

1506400 

DIGITAL METER 

MOT1 

1560715 

MOTOR 

MOV7 

1606100 

MOVISTOR 

PS1 

1590103 

LINEAR POWER SUPPLY 

RIBBON CBL 

1079910 

RIBBON CABLE 

RIBBON CBL 

1150191 

RIBBON CONNECTOR 

SW1 

1863037 

ROTARY SWITCH 

SW2 

ROCKER SWITCH, 120V, 5A, MOM-OFF-MOM, SPDT 

SW3-6 

1866015 

LIMIT SWITCH 

SW7 

ROCKER SWITCH, 120V, 5A, MOM-OFF-MOM, DPDT 

SW8 

ROCKER SWITCH, 120V, 5A, ON-MOM, SPDT 

T3 

1890105 

VARIABLE TRANSFORMER 

 

7010993797 

PCB1402: LD60A WITH BLUETOOTH 

 

38322205 

GA1-2205 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for LD60A

Page 1: ...LD60A User s Manual LIQUID DIELECTRIC TEST SET MODEL NUMBER LD60A VERSION 3 4 Copyright Phenix Technologies Inc Rev 2 28 2018 Rev 8 26 2019 format...

Page 2: ...ONS DESCRIPTION 1 SPECIFICATIONS 2 INSTALLATION AND GROUNDING INSTRUCTIONS 3 OPERATING INSTRUCTIONS 4 CALIBRATION 5 ELECTRICAL SCHEMATICS 6 MAINTENANCE AND TROUBLESHOOTING 7 PARTS LIST 8 PARTS ORDERIN...

Page 3: ...LD60A DANGER WARNINGS WARNING Complete Grounding of this unit is necessary for the safe operation of this equipment Disconnect inputs before ungrounding this equipment...

Page 4: ...t have been listed below for your reference Become familiar with your instrument before performing an actual test Know your work area check that all circuits are de energized and locked out Never work...

Page 5: ...lat electrodes according to ASTM D877 Other cells are available upon request SAFETY FEATURES Center tapped high voltage transformer The test cell is isolated from the supply when the protective lid is...

Page 6: ...t Voltage Power 0 60 000 Volts at 500 VA momentary Maximum Voltage to Earth is 30 000 Volts Rate of Rise Selectable 500 2000 3000 Volts Second Memory Voltmeter Scale of 0 60 000 Volts Accuracy 1 of Fu...

Page 7: ...ilar flat and stable surface 2 Make sure that the test set is level 3 Open the cover 4 Check that your input voltage corresponds with the input of the test set 5 Check that your supply outlet provides...

Page 8: ...ates if test fails 4 Press to Start or to Stop 5 Remote Mode Press to select OFF or ON If ON controls are available on the optional Tablet or Bluetooth device if enabled See instructions below 6 Devic...

Page 9: ...est compartment 2 Set voltage rate of rise switch to correct position 3 Close test chamber lid 4 Switch to start position It will automatically raise the voltage to the breakdown level indicate failur...

Page 10: ...be sent to a printer or shared via email Pairing with the LD60A If for some reason you need to re establish the paring with the LD60A follow the instructions in this section Pairing creates a bond be...

Page 11: ...The device is powered and ready Blinks continuously Connected to host The BT Power LED blinks alternately with the Pair LED the device is in pairing mode Off Device is off Pair LED Indication Steady...

Page 12: ...sample of oil according to standards procedure 3 Place the test cell in the test chamber of the unit and close chamber lid 4 Turn on AC power switch 5 Select 3000 volts per second rate of rise 6 Move...

Page 13: ...ith test sample of oil according to standards procedure 3 Place the test cell in the test chamber of the unit 4 Plug cord of stirrer into chamber receptacle 5 Close chamber cover 6 Turn on AC power sw...

Page 14: ...speed 5 Move the Start Stop switch to the Start position 6 High voltage will come on the unit will automatically raise the voltage at the selected rate per second 7 Allow voltage to rise on the LDCU...

Page 15: ...sary For any internal changes modifications or calibrations found necessary please return to the factory Calibration of LDCU If your unit includes the optional LDCU unit it should be sent to Phenix Te...

Page 16: ...LD60A 6 1 ELECTRICAL SCHEMATICS Drawing Number Description 1 9101060 LD60A LIQUID DIELECTRIC TEST SET...

Page 17: ...wer source CAUTION Never attempt to clean inside the unit as the cleaning solution may cause damage to the electronic components In the event it becomes necessary to replace any parts a description ca...

Page 18: ...ELLOW 1 LED3 4 LED GREEN 2 METER 1506400 DIGITAL METER 1 MOT1 1560715 MOTOR 1 MOV7 1606100 MOVISTOR 1 PS1 1590103 LINEAR POWER SUPPLY 1 RIBBON CBL 1079910 RIBBON CABLE 3 RIBBON CBL 1150191 RIBBON CONN...

Page 19: ...include the serial number of the unit as well as the part number of the replacement part When your purchase order is received at our office a representative of Phenix Technologies will contact you to...

Page 20: ...l Number Reason for Return Cause of Defect If Phenix Technologies Inc deems return of the part appropriate it will then issue an Authorization for Return If return is not deemed advisable other inspec...

Page 21: ...or s Manual is correct If there are concerns or comments as you have used this information Phenix Technologies appreciates any feedback Unit Serial Number SECT PAGE S COMMENT Please return to Phenix T...