CHAMBER TEMPERATURE SETTING
Set the chamber temperature according to the condition of use. This refrigerator can keep the stored
items for long period under appropriate temperature.
Setting range of chamber temperature: between 2
o
C and 14
o
C
Initial setting (factory setting): 5
o
C
<Important>
The chamber temperature of 2
o
C may cause partial freeze of stored items.
Example: Change the chamber temperature to 4
o
C from 5
o
C
►
Following shows a sample setting. Set the desired chamber temperature according to the condition
of use.
Description of operation
Key operated
Indication after operation
1
Connect the power supply plug to
the dedicated outlet.
(only when start-up of unit)
-----
The current chamber temperature is
displayed.
2 Press set key.
SET
The current setting (005) is displayed
and the first digit blinks.
3
Press up arrow key and scroll the
figure to 4.
The display is changed to 004 from
005.
4 Press set key.
SET
Set temperature is memorized and
the current chamber temperature is
displayed.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function). In this case, the setting is not accepted.