background image

42 

 

VARIOUS SETTING (MENU/Tools) 

 

   

Default setting (Tools/Default Setting) 

 
1.

 Select “Default Setting” in the Select Tools screen.    Press the menu button (MENU) to show the menu 

window.    Select “OK”, and press the enter key (ENTER).

   

 

 

 
 
 
 
 
 

2.

 The Default Setting screen is displayed.    Set the default for each parameter as necessary.

 

 
 
 
 
 
 
 
 
 

LCD Back Color:

    Setting of background color (1. Blue    2. White) 

Buzzer Finished:

    Select of buzzer activation (1: Yes) or no activation (2: No) at the time of completion 

of a programmed running. (The buzzer activates 6 times when a program is finished.) 

Analog Range:

    Select of analog output range of temperature in chamber (0: 0

 o

C~50

 o

C  1: -20

 o

C~80

 

o

C) 

 
Note: 

DAQ is an external monitoring system of chamber status.    It is necessary to set the DAQ Speed, DAQ ID 
and DAQ Mode when using communication software.  Communication software is ordered specially 
contact our sales representative or agent.

 

 

 
 

 

                                                                               

 

    D e  f  a  u  l  t    S e  t  t  i  n g                                              

 

    L C D   B a  c  k    C o l  o r     

 ( 1 

. B 

l u 

e  2 . Wh 

i t e )     

 

  D 

 S 

e d 

       0 

 ( 0 

. 2 

 2 

. 9 

0 0 )      

 

    D A Q   I  D                     0    (  0  .  O F F      1  -  2  5  0  )       

   

 

 

    D A Q   M o d e                  0    (  0  .  L o c  a  l    1  .  R e  m o  t  e  )     

 

  B 

z z e r  F 

i n 

i s 

 1 

 ( 1 

. Y 

 2 

. N 

)         

 

  A 

l o 

 R 

    0 

 ( 0 

. N 

r m 

l  1 

. Wi d e )      

1  2  3  4  5  6  7  8  9  0  1  2  3  4  5  6  7  8  9  0  1  2  3  4  5  6  7  8  9  0  1  2  3  4  5  6  7  8  9  0 

  S 

e l e 

c t  T 

l s   M 

I R 

- 1 5 4 

                 

   

L o g  

       

                                                           

  D 

a t e 

 T 

i me                                                           

    A l a r m , O p t . , L o w   H u mi .   S e t t i n g      

           

 

    K e y   L o c k   P W  S e  t  t  i  n  g                                      

 

 

D e  f  a  u  l  t    S e  t  t  i  n  g                           

                   

    D e  l  e  t  e    U s  e  r    D a  t  a                                             

    M E N U    

  O K          

 S 

v c     

  C a n c e l   

Summary of Contents for MIR-154 Series

Page 1: ...54 Series Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 61 for model number Operating Instructions Cooled Incub...

Page 2: ......

Page 3: ...PROGRAMMING MENU Edit 22 EDIT OF SAVED PROGRAM MENU Edit 28 START OF PROGRAM MENU Run 29 JOIN FUNCTION 31 SKIP OF STEP MENU Skip 32 STOP OF PROGRAM MENU Stop 33 AUTOMATIC DEFROST MENU Tools Date Time...

Page 4: ...2 CONTENTS TROUBLESHOOTING 51 DISPOSAL OF UNIT 52 LOW HUMIDITY MODE 57 LOW COMPRESSOR VIBRATION DESIGN 58 PERFORMANCE DATA 59 SPECIFICATIONS 60 PERFORMANCE 61 SAFETY CHECK SHEET 62...

Page 5: ...t our sales representative or agent if any page of the operating instructions is lost or the page order is incorrect Contact our sales representative or agent if any point in the operating instruction...

Page 6: ...ersonnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Symbol shows This sym...

Page 7: ...d safe disposal options visit ec gc ca mercure mercury Dispose of or recycle in accordance with applicable laws Pour plus de renseignements sur les proc dures de manutention s curitaire les mesures pr...

Page 8: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Page 9: ...fire When removing the power supply plug from the power supply outlet grip the power supply plug not the cord Pulling the power supply cord may result in electric shock or fire by short circuit Discon...

Page 10: ...as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components or electric parts Check the setting when starting up of operation after power failure or...

Page 11: ...C Maximum relative humidity 80 for temperature up to 31o C decreasing linearly to 50 relative humidity at 40o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltages...

Page 12: ...10 INCUBATOR COMPONENTS 1 MIR 154 2 4 3 5 7 6 8 9 10 11 12 13 14 1 2 3 4 5 6 7 8 9 10 11 12 13 14 16 MIR 254 15 15...

Page 13: ...ber stopping to the hole for measurement placed at the left side as it was when passing cables for measurement and power cords through it Be sure to replace the cap after take out cable or the inside...

Page 14: ...silence the alarm buzzer temporarily 7 Character input key 8 Clear key CE To clear the input value during editing of program 9 Enter key ENTER To determine the selection of menu In program editing pre...

Page 15: ...put Permissible contact capacity DC 30 V 2 A a Contact output Connect the lead wires to COM and NO terminals Normal condition open Abnormal condition close b Contact output Connect the lead wires to C...

Page 16: ...e a malfunction It is due to the hot gas piped around the unit frame to prevent condensation around the cabinet A location with a sturdy and level floor Always install the unit on a sturdy and level f...

Page 17: ...ution refer to the instruction of the detergent After the cleaning with the diluted detergent always wipe it off with a wet cloth Then wipe off the panels with a dry cloth Note Remove the cable tie ba...

Page 18: ...ge display field A message is displayed when a breakdown occurs 6 Alarm display Alarm Alarm is alternately displayed by reversed non reversed character while alarm is operating Warning is alternately...

Page 19: ...installed page 40 Setting of low humidity mode To set the running for low humidity page 40 Default setting To set the default for LCD panel and communication DAQ speed etc page 42 STANDBY OPERATION M...

Page 20: ...be changed Setting of Key Unlock Key Unlock 1 When setting of key unlock change the value of the key lock line from 1 to 0 in the stand by setting screen Stand by Setting and press the enter key ENTER...

Page 21: ...racter input key Chamber temperature setting value 10 0 is displayed 6 Press the menu button MENU MENU MENU window is displayed 7 Select in MENU window pressing the shift key downward is displayed in...

Page 22: ...er key ENTER ENTER The set value is memorized and top screen is displayed Procedures of Key Unlock Changing Key Lock setting from 1 Lock to 0 Unlock Operation Key LCD Display 1 Top screen is displayed...

Page 23: ...imit temperature alarm Low Limit 5 o C or lower than the minimum temperature in a program or stand by operation 5 Press the menu button MENU at the completion of setting The menu window is shown Selec...

Page 24: ...ect Edit and push the enter key ENTER 3 The Program Name Stored screen is opened Press the menu button MENU and select New and press the enter key ENTER The program names are displayed when some progr...

Page 25: ...hanged by Insert or Delete on the menu window Press the menu button MENU to open the menu window The menu window for a first section has no Insert or Delete Therefore neither insert nor delete is effe...

Page 26: ...ture is between 10 0 o C and 60 0 o C The temperature control range is between 2 o C and 50 o C when light add on kit MIR L15 sold separately is installed In this case temperature fluctuation may exce...

Page 27: ...00 15 00 17 00 Li ght 0 0 0 0 0 0 LS Time 19 00 21 00 22 00 23 00 Li ght 0 0 0 0 When setting Light step LS displayed at the upper left and Light at the lower left on the top screen 1 0St p 31Cyc LS T...

Page 28: ...saved Edit function of characters Shift key Upward shift key Space insertion Downward shift key backspace Leftward shift key Move a cursor left Rightward shift key Move a cursor right Character input...

Page 29: ...s displayed and 99 is changed into Display of changes 99 when lapped by a cursor 2 Set a time temperature and light step as same as Example 1 3 Input a program name NIKKO pres the menu button MENU to...

Page 30: ...menu window Select Save when saving by overwriting or select SaveAs when saving with another program name Press the enter key ENTER The edited program is entered Do not select SaveAs with same program...

Page 31: ...K and press the enter key ENTER Timer selection of Clock mode or Timer mode 1 Clock Clock mode Displays start time of each steps 2 Dec Timer mode Displays the remaining time up to a next step Join Joi...

Page 32: ...utton MENU to show the menu window Select Graph and press the enter key ENTER When the starting time of program operation is later Waiting is displayed in the current running status of LCD panel O z e...

Page 33: ...ut NIKKO as the same character string the input for the join function is as follows NIKKO 1 Autumn NIKKO 2 Spring NIKKO 3 Summer 2 When running the joined program in Ex 1 select the program Oze 1 Spri...

Page 34: ...menu button MENU Selecting Next causes the skip to the next step Selecting Back causes the skip to the previous step After selecting Next or Back press the enter key ENTER 4 Programmed running shifts...

Page 35: ...and press the enter key ENTER 3 The Confirmation screen is displayed Select OK and press the enter key ENTER to stop the program 4 After stopping the program the top screen is displayed C o n f i r m...

Page 36: ...CAUTION During defrosting Defrosting is displayed on running status of LCD panel Care should be taken for deviation of setting and current temperature during defrosting In case of long term running o...

Page 37: ...osting Defrosting is displayed at the upper left on the top screen CAUTION The manual defrost can be started during programmed running standby operation or automatic defrosting 4 The manual defrost is...

Page 38: ...shift key older date rightward shift key newer date S t a n d b y T o p S c r e e n T e mp 2 5 0 o C A l a r m N o r ma l D o o r C l o s e d L i g h t O F F 2 0 0 8 0 6 2 0 1 2 0 0 0 0 M E N U R u n...

Page 39: ...10 o C Similarly select Range1 and press the enter key ENTER Upper limit is changed from 60 o C to 50 o C and lower limit is changed from 10 o C to 0 o C 6 0 o C 2 0 0 8 0 6 2 0 0 1 0 o C 1 2 2 4 5 0...

Page 40: ...ation of hyper terminal on PC Apply txt or csy to retention file as an extension Press the menu button MENU to show the menu window Select Start and press the enter key ENTER The transmission is start...

Page 41: ...val 6 minutes About 14 days 3 Log interval 30 minutes About 70 days After passing the memory limit the older data is deleted and newer data is memorized Setting of automatic defrost Select one of two...

Page 42: ...outside the 2 o C to 50 o C range when the program already saved is running The set value of temperature is limited between 2 o C 50 o C after Lamp is set to 1 4 The condensation in the chamber is re...

Page 43: ...4 Input User Password 4 digits again and press the enter key ENTER K e y L o c k P W S e t t i n g C u r r e n t U s e r P a s s w o r d K e y L o c k P W S e t t i n g N e w U s e r P a s s w o r d...

Page 44: ...o C 80 o C Note DAQ is an external monitoring system of chamber status It is necessary to set the DAQ Speed DAQ ID and DAQ Mode when using communication software Communication software is ordered spec...

Page 45: ...and press the enter key ENTER The program Oze 1 Spring is now deleted D e l e t e U s e r D a t a D e l e t e a U s e r P r o g r a m D e l e t e A l l U s e r P r o g r a ms C o n f i r m a t i o n...

Page 46: ...y ENTER 6 The Confirmation screen is displayed The menu window is already displayed Select OK and press the enter key ENTER All of the programs are now deleted C o n f i r m a t i o n A r e y o u s u...

Page 47: ...nsure that the intake and exhaust vents are not blocked Adequate space should be provided between the items inside the unit to allow air circulation And always install the stainless plate on the botto...

Page 48: ...heater and sub heater OFF Thermal sensor abnormality Input voltage is over 70 o C corresponding Input voltage is lower than 50 o C corresponding Top screen E01 Temp sensor is opened E02 Temp sensor i...

Page 49: ...orresponds to 50 o C or less disconnection of sensor Top screen E09 Comp sensor is opened Intermittent tone Alarm status Compressor OFF Defrost sensor abnormality The input voltage corresponds to 50 o...

Page 50: ...f under the program operation The alarm is activated as follows under program operation The buzzer is stopped by pressing the alarm buzzer stop key BUZZER but the remote alarm keeps alarm status The b...

Page 51: ...onto or into the unit Doing so can damage the electrical insulation and may cause electric shock or short circuit The compressor and other mechanical part are completely sealed This unit requires abso...

Page 52: ...n the both sides remove the socket from the lamp 4 Attach the socket to a new lamp and then replace the water proof rubber 5 Replace the shade and fix with 2 screws The lamp is consumable Replacement...

Page 53: ...al temperature of chamber reaches to the set temperature of operation The excessive heat source is in the chamber Remove the excessive heat source For the allowable heat load refer to the graphs on pa...

Page 54: ...ended period ensure that children do not have access and doors cannot be closed completely The disposal of the unit should be accomplished by appropriate personnel Always remove doors to prevent accid...

Page 55: ...cal involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitdokumenten...

Page 56: ...r sur les piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci...

Page 57: ...so estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazioni con s...

Page 58: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Page 59: ...e material in the chamber may be dried depending on the usage condition when the low humidity mode is selected Do not select the low humidity mode when the drying of culture media should be avoided Wh...

Page 60: ...n you want to reduce the vibration Stick the caution label packed with this product when low compressor vibration design is applied The refrigeration circuit is at serious risk of damage if the produc...

Page 61: ...r Maximum heat load tolerable in chamber Heat load input in chamber Ambient temperature 1 Temperature reached in chamber 10o C 2 Temperature reached in chamber 0o C 3 Temperature reached in chamber 10...

Page 62: ...foamed in place Cooling method Forced air circulation Compressor Reciprocated compressor Output 150 W Output 250 W Evaporator Fin and tube type Condenser Wire and tube type Refrigerant R 134a R 404A...

Page 63: ...ient temp 5 o C 35 o C no load Temperature fluctuation 1 5 o C ON OFF control set 5 o C ambient temp 20 o C no load 0 2 o C PID control set 50 o C ambient temp 20 o C no load Temperature uniformity 0...

Page 64: ...f the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see be...

Page 65: ......

Page 66: ......

Page 67: ......

Page 68: ...Printed in Japan LDCL043000 0 S0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Reviews: