31
SPECIFICATIONS
Product name
Blood Bank Refrigerator
MBR-305GR
External dimensions
W600 mm x D680 mm x H1835 mm
Internal dimensions
W520 mm x D490 mm x H1150 mm
Effective capacity
302 L
Exterior Painted
steel
Interior Painted
steel
Outer door
Painted steel, Provided with 2-layer glass window
Inner door
Acrylic resin, 2 doors
Insulation
Rigid polyurethane foamed-in place
Drawer
Stainless steel x 5, Allowable load; 20 kg/drawer
Access port
Inner diameter; 30 mm, Left side
Cooling method
Forced cool air circulation
Compressor
Hermetic type, Output; 150 W x 1
Evaporator
Fin and tube type
Condenser
Wire and tube type
Refrigerant R-134a
Defrosting
Cycle d forced defrost, Fully automatic
Defrost heater
50 W
Temperature controller
Electronic control system
Temperature display
Digital display (in 0.1
o
C increments)
Fluorescent lamp
FL15SD x 1
Alarms
High temp. alarm, Low temp. alarm, Power failure alarm, Door alarm
Battery
For power failure alarm (DC 6 V 1100 mAh, automatic recharge)
Remote alarm contact
During alarm buzzer and power failure, Output; N.O.; “close”, N.C. “open”
Weight 147
kg
Accessories
1 set of key
Optional components
Interface board (MTR-480)*
Design or specifications will be subject to change without notice.
Refer to the updated catalog when ordering an optional component.
* For the data acquisition system MTR-5000 user only.