Phcbi MBR-305GR Operating Instructions Manual Download Page 1

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

MBR-305GR

Series

Please read the operating instructions carefully before using this product, and keep the operating 

instructions for future use. 

See page 32 for all model numbers.

 

Operating Instructions

 

Blood Bank Refrigerator 

Summary of Contents for MBR-305GR

Page 1: ...ries Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 32 for all model numbers Operating Instructions Blood Bank R...

Page 2: ......

Page 3: ...R TEMPERATURE P 16 DEFROSTING OF EVAPORATOR P 16 REMOTE ALARM TERMINAL P 17 CALIBRATION OF SENSOR P 18 ALARM FUNCTIONS P 19 SELF DIAGNOSTIC FUNCTIONS P 20 ROUTINE MAINTENANCE Cleaning of exterior inte...

Page 4: ...provement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page order is incorrect Contact our sales representative or agent i...

Page 5: ...t in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Sym...

Page 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Page 7: ...al operation may cause electric shock or fire When removing the plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short...

Page 8: ...r improper insertion may cause a heat or ignition Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components o...

Page 9: ...C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltag...

Page 10: ...8 REFRIGERATOR COMPONENTS Refrigerator unit 7 1 Florescent lamp page 22 Temperature recorder Light switch 2 3 4 5 6 8 10 11 9 Glow starter page 23 Power switch Door gasket 9...

Page 11: ...ent bottom front Do not block this vent Blocking this vent may cause unstable chamber temperature Arrange the storage items not to subject to the cold air from this vent 7 Leveling feet front 2 locati...

Page 12: ...n error code and a chamber temperature is displayed alternately when the self diagnostic function detects any abnormality page 20 6 Up arrow key This key is used at calibration mode of sensor for temp...

Page 13: ...stall the unit on a sturdy and level floor withstanding the total weight of the unit The uneven floor or tilted installation may cause failure or injury Unstable condition may cause vibration or noise...

Page 14: ...ounterclockwise to separate 2 front casters from the floor Fig 1 And adjust the leveling foot to level the unit The unit is installed firmly by separating the front casters from the floor The unit may...

Page 15: ...e than 2 minutes The alarm buzzer is canceled when the outer door is closed 3 The chamber temperature is set to 4 o C at the factory Check that the chamber temperature reaches 4 o C on the temperature...

Page 16: ...xhaust by the storage items near the air exhaust vent Care should be taken not to put storage items nearby the air exhaust vent The storage items may be frozen Load line The storage area is limited to...

Page 17: ...herefore it is removed together with the cover Take out the recorder sensor as well from the lower monitor bottle 4 Remove the fixing screws 2 screws and take out the monitor bottle 5 Remove the cap s...

Page 18: ...temperature display changeover key Lower temperature is displayed UPPER LOWER Lower temperature 4 Press the temperature display changeover key Mean temperature is displayed UPPER LOWER Mean temperatu...

Page 19: ...is canceled by pressing the alarm buzzer stop key BUZZER since the remote alarm terminal is operated in conjunction with alarm buzzer However the alarm status of remote alarm terminal is not canceled...

Page 20: ...ting the mixture allow the temperature indicated on the temperature display to be stable 7 Calibration is needed if the temperature on the temperature display does not show 0o C 8 Press the calibratio...

Page 21: ...However the alarm status of remote alarm terminal is not canceled by pressing the alarm buzzer stop key BUZZER when the unit is under power failure alarm This unit is provided with the alarm buzzer re...

Page 22: ...and chamber temp is displayed alternately If the upper sensor for temp display is short circuited Alarm indicator blinks E16 and chamber temp is displayed alternately If the lower sensor for temp disp...

Page 23: ...t cloth Then wipe off the cabinet or accessories with a dry cloth Important Never pour water onto or into the unit This may cause electric shock or failure CAUTION Do not use a brush an acid a thinner...

Page 24: ...e clip 2 locations together with the wiring Fig 1 4 After removing the water proof cover remove the socket with the wiring on the both sides of the lamp cover Fig 2 5 Take out the fluorescent lamp fro...

Page 25: ...Turn off the light switch and power switch and then disconnect the power supply plug 2 Remove the glow starter and install a new glow starter glow starter FG 1P 3 Connect the power supply plug to the...

Page 26: ...lacement of the battery for power failure alarm involves the risk of electric shock Important The used battery is a recyclable precious resource Do not dispose of the battery Always follow the procedu...

Page 27: ...breaker on the supply circuit is activated A breaker of 15 A or more is recommended The fuse on the supply circuit is blown The outer door was kept opened for a long time The outer door is opened An e...

Page 28: ...ersonnel Always remove doors to prevent accidents such as suffocation Recycle of battery The unit contains a rechargeable battery The battery is recyclable At the end of its useful life check with you...

Page 29: ...the chemical involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitd...

Page 30: ...r sur les piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci...

Page 31: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Page 32: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Page 33: ...r Fin and tube type Condenser Wire and tube type Refrigerant R 134a Defrosting Cycle defrost forced defrost Fully automatic Defrost heater 50 W Temperature controller Electronic control system Tempera...

Page 34: ...o C to 35 o C Noise level 41 41 dB A scale 50 Hz 60 Hz Maximum pressure 1850 kPa Rated voltage AC 115 V AC 220 V AC 220 V Rated frequency 60 Hz 60 Hz 60 Hz Amperage 3 0 A 1 7 A 1 5 A Power consumption...

Page 35: ...Contamination of the unit Unit interior No contamination Yes No Decontaminated Yes No Contaminated Yes No Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to...

Page 36: ...ginal Operating Instructions Printed in Japan LDCL038300 0 S0418 0 2018 04 01 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018 Nijverheidsweg 120 4879 AZ Etten Leur The Neth...

Reviews: