Phase Technology 70Xi-APA/ APC/ ASL Analyzer System – Quick Installation Guide

 

Page 1 of 3 

Customer Service & Support:   604-241-9568 

 

 

[email protected] 

 

1.

 

Remove the following items from the 3 boxes: 

70Xi analyzer box: 

 

1 70Xi analyzer 

 

1 accessory box: 

1 AC power 
cable 

1 dryer bottle 

 

 Autosampler 

box: 

 

1 Autosampler 

 

1 accessory box: 

1 AC power cable 

1 RS232 cable 

2 vial heater/sample 
rack 

(*APA & APC: 2 vial heaters) 
(*ASL: 1 vial heater & 1 rack) 

Accessory box: 

 

1 waste tank with grounding cable 

 

1 box with solvent tank with grounding cable 

 

1 box with pour sprout (level sensor attached) 

 

1 bag of tubings and cables 

 
 

2.

 

Set up the analyzer system: 

 

The system should be set up: 

on a flat, vibration-free bench with room underneath 

in a well-ventilated location 

 

Put the autosampler on the right side of the analyzer, with 
the AC power cable connector facing the back.  

 

Put the waste tank and solvent tank under the bench. 
 

 

 
 
3.

 

Remove 2 plastic caps and 7 swage caps on the back of the 
analyzer, and 3 plastic caps at the solvent tank.  Keep them 
with the shipping boxes and related packing materials. 

 

 

4.

 

Connect the dryer bottle to the back of the analyzer: 

a)

 

Attach the dryer bottle to the 
back of the analyzer by fitting 
the two mounting bracket holes 
over the two mounting screws.  

 

b)

 

Separate the quick-connect 
fittings on the dryer bottle. Press 
the latch on the female fitting 
and pull them apart.  

 

c)

 

Attach the quick-connect fittings 
to the “PURGE IN” and “PURGE 
OUT” fittings on the back of the 
analyzer. A “click” sound 
confirms that the fittings have 
snapped into place. 

 
 
 

5.

 

Load the two vial heater/sample rack to the autosampler.  
Make sure the 4 feet of each vial heater/sample rack sit 
inside the square cutouts of the autosampler bottom tray. 

 
 
 

 

8 square cutouts on the 
bottom tray 

 

 

Vial heater/sample 
rack on the bottom 
tray (*put the vial 
heater with wash pot 
at front) 

 

 

6.

 

Use wrench to connect the tubing to back of the analyzer: 

Port at back of analyzer: 

Tubing: 

“From Base of Wash Pot” 

 

 

Labeled 

“From Base of Wash Pot” 

“To Waste Tank” 

 

 

Labeled 

“To Waste Tank” 

“Air to Solvent Tank” 

 

 

Labeled 

“Air To Solvent Tank” 

“Sample From 

Auto-Sampler” 

 

 

Labeled 

“Sample From Autosampler” 

(the other end of the tubing 

already attached to the 

larger needle of the 
autosampler probe) 

“Solvent From 

Solvent Tank” 

 

 

Labeled “Solvent From 

Solvent Tank” 

“From Top of Wash Pot” 

 

 

 

Labeled 

“From Top of Wash Pot” 

(the other end of the tubing 

already attached to top of 

wash pot of the vial heater) 

 

“Air to Auto-Sampler” 

 

 

Labeled 

“Air To Autosampler” 

(the other end of the tubing 

already attached to the 

smaller needle of the 

autosampler probe) 