pg instruments T60U User Manual Download Page 29

24

6.2 Maintenance 

1.  The screen fails to display or doesn’t display clearly after power on 

Check the contrast dial--use the plain screwdriver to adjust the “CONTRAST” dial rear of the 

instrument.

2.  The instrument does not operate at all; noting appears on LCD screen. 

Check the mains voltage according to the page five “check the mains voltage”. 
Check the power cord connection and connect the power cord properly. 
Check whether the fuse is blown. If it is, replace it.

3.  The printer does not operate; printout failure 

Check whether the types of printer are right. T60 only supports HP ink jet printer (HP 5652 is 

recommended) and mini temperature sensing parallel printer (WH153PT01 is recommended). 

Check whether the printer connection is loose 
Check whether the printer selection is right. UP stands for mini temperature sensing parallel printer, 

while HP stands for HP inkjet parallel printer or laser jet parallel printer.

4.  “ Cell init ERR” of initialization fails 

Check whether there is opaque object in the sample compartment block the motor movement.
5.

“Lamp motor”, “W lamp”, “D2 lamp” and “WL check” of initialization fail

 

Check whether there is opaque object or cell in the sample compartment. Remove the opaque object. 

Take out the cell after each measurement to avoid unnecessary intervention and corrosion to the instrument. 

Check whether the mains voltage is too low, which may cause abnormal startup. 
Check whether the W lamp or the D2 lamp is lit; if not, replace it. 

6. Unstable 

measurement  

Check whether the instrument performs initialization normally. If you press 

Ǐ

RETURN

ǐ

 soon after 

you power on, the instrument will not perform initialization or operate normally.   

Take out the cell at current measurement status, and then press 

Ǐ

ZERO

ǐ

  to check whether 0.000Abs 

is fluctuate. If it fluctuates between –0.001 and +0.001, it is normal. If it fluctuates widely, please confirm 
the outer power supply meets the power requirements. Then, restart the instrument to perform initialization, 
and confirm the initialization is normal. 

If the sample compartment is blank, press 

Ǐ

ZERO

ǐ

 to check whether the 0.000Abs fluctuates 

normally. Check whether the following problems exist: 

1)

Check whether the volatility of the measured sample is high. If it is, please use the cell cover. 
If you use the highly volatile gas, such as benzene, please perform measurement after remove 
the interference gas by opening the sample compartment. 

2)

Check whether the reagent blank used to zero meets the requirements. It stipulates that the 
Abs of the reagent blank should be no more than 0.4Abs. If it is out of range, please replace 
the reagent blank or the reference blank.   

 

7.  Poor reproducibility of the measured sample 

Check whether the sample is stable and whether there is sample photolysis. 
Check whether you use the right way to wipe off the cell. The cell should be wipe and clean carefully.

8.  The card program is inoperative 

Check whether you insert the program card. Make sure you insert it correctly. 
Please refer to the instruction manual of the program card for the right way of loading it.

9.  The Abs is not correct 

Perform “Dark current correction”, then perform reagent blank, and then remeasure. 
Confirm that the cells match each other.

10. Solutions to the frequently asked questions: 

How to perform 0%T correction? 

T60 is an automatic instrument, which should not perform 0%T correction each time when measuring. 

For the automatic instrument, 0%T correction is also called dark current correction, which should not 
always be performed. When the environment changes (e.g. temperature, voltage, ambient brightness), 0%T 
correction needs to be performed. Select “dark current correction” from system set, and then performs. 

How to change the tungsten lamp from the deuterium lamp, and vice versa? 

Summary of Contents for T60U

Page 1: ...instruments Analytical Instruments for Science U S E R M A N U A L T60U U V V I S S P E C T R O P H O T O M E T E R...

Page 2: ...ments of the analyzer For obtaining maximum performance and trouble free operation we give you the following suggestions Upon receiving your instrument please check all the items listed in the packing...

Page 3: ...d dismount the instrument Caution Improper operation may cause unnecessary damage or accidents Eyes protection Please wear glasses for the inner ultraviolet radiation and strong light Scald Don t touc...

Page 4: ...INSTRUMENT CONSTRUCTION 6 2 1 Front view 6 2 2 Side view 6 2 3 Rear view 7 2 4 keypad description 7 CHAPTER 3 OPERATION 8 3 1 Introduction 8 3 2 Photometric Measurement 9 3 3 Card Program 15 3 4 Syste...

Page 5: ...ENTS LIMITED Information in this publication is subject to change without notice as PG INSTRUMENTS LIMITED continues to develop the latest technologies and products for our customers The units describ...

Page 6: ...sent Serial number Name No Q ty 01 T60 main unit 1 02 Conformity certificate 1 03 Fuse T 250V 2A 2 04 Quantitative program card 1 05 Installation guide 1 06 Black block 1 07 AC power cord 1 08 Instruc...

Page 7: ...chlorine hydrochloric acid gas hydrogen sulphides gas sulfurous acid gas 5 Area with very little dust or other airborne particles 6 Placing on the stable workbench The workbench should be stable and...

Page 8: ...wer or it may damage the instrument T60 is designed to operate on 230V 200VA and 50 60Hz mains supply The original setup voltage for T60 is 230V Confirm the AC mains is earthed in the building before...

Page 9: ...ibration Step two Connect to the printer Be sure to turn off the power of the instrument and the printer before connecting via the parallel port otherwise it may damage the devices Step three Connect...

Page 10: ...5 Step five When the initialization procedure is completed the main menu will automatically appear on the screen...

Page 11: ...6 Chapter 2 Instrument construction 2 1 Front view 2 2 Side view...

Page 12: ...Set the working mode Set the wavelength Zero Abs 100 T Print data Edit keys Start measurement Used to quit operation and return Used to confirm the current parameters and operating mode Numeric keys...

Page 13: ...at a fixed wavelength and K factor and print out Card Program The instrument can perform a required measurement such as quantitative measurement DNA Protein analysis etc by loading a pre programmed ca...

Page 14: ...nt Fig 3 2 press SET to gain access to the menu of parameters setting Fig 3 3 Press and to move the cursor to the required option then press ENTER to gain access to the required menu Press RETURN to q...

Page 15: ...u as shown in Fig 3 6 Fig 3 6 At the bottom of this page enter the K factor by pressing the numeric keys 0 9 decimal point and negative sign Press RETURN key to clear the mistakes and press CE to clea...

Page 16: ...1 Reagent Blank Press and to move the cursor to the required option then press ENTER to toggle between yes and no If you choose yes the system will take the sample in the No 1 Cell as the reagent blan...

Page 17: ...ell Abs tn The blank value of NO n sample cell transmittance Move Cell Press and to move the cursor to the required option then press ENTER The sample cell will move forward one cell by pressing ENTER...

Page 18: ...a new entry again Auto Zero In the case of photometric measurement Fig 3 1 press ZERO to zero the Abs or 100 T at current wavelength Put the blank sample into the sample compartment before correction...

Page 19: ...rting measurement the next time The user can select print or not If the user chooses Print the system will print the measured data If the user chooses Don t print the system will clear the data area C...

Page 20: ...Unloading the program card Please turn off the instrument before taking out the card Press the pushing bar above the slot to take out the card Inputting the serial number When the user uses the progr...

Page 21: ...tely It costs 20 seconds for the deuterium lamp to change from off to on The system will display BUSY and neglect the user s input during this period Tungsten Lamp In the case of system set press and...

Page 22: ...me RETURN key is used to clear mistakes and press CE key to clear step by step then press ENTER key to confirm If nothing is to be entered press RETURN key to quit this page directly A 2 digit figure...

Page 23: ...to the WL Correct then press ENTER The system will test and display the current characteristic wavelength PEAK n and prompt you whether to perform wavelength correction When the absolute value of n 65...

Page 24: ...ecommended to reset system Input Change Lamp Wavelength In the case of system set press and to move the cursor to the Change Lamp WL then press ENTER to input value Fig 3 21 At the bottom of this page...

Page 25: ...20 About In the case of system set press and to move the cursor to the option of About then press ENTER Fig 3 22 Press RETURN to return to the menu of system set...

Page 26: ...01A 1h 500nm after 1 hour warm up Photometric Range Abs 0 3 3Abs Transmittance 0 200 T Interface PC interface via RS232 link Printer interface via parallel output port Optional Program Card With the e...

Page 27: ...and user friendly software can greatly improve your work efficiency The functions Photometric measurement Quantitative measurement Spectrum scanning Kinetics measurement 3D presentation The software...

Page 28: ...r There are four dust filters at the bottom of the instrument Generally speaking the user should clean them every three months But if in the places under severe conditions such as with much sand it sh...

Page 29: ...it fluctuates between 0 001 and 0 001 it is normal If it fluctuates widely please confirm the outer power supply meets the power requirements Then restart the instrument to perform initialization and...

Page 30: ...ion if the instrument is placed in the humid environment for a long time it will also cause the stray light raises higher It is recommended that switch on the power of the instrument once a week to av...

Page 31: ...bottom counterclockwise to make it vertical to the cover as shown in Fig 7 1 Fig 7 1 Step two loosen the two knobs at the right rear of the instrument Fig 7 2 Step three take off the right cover Fig 7...

Page 32: ...pair of cotton gloves Screw up the two crews Don t touch the light source cover if the instrument has worked for a while and the cover is hot The ultraviolet radiation emitted by the deuterium lamp a...

Page 33: ...28 Appendix 2 Clean the dust filter Switch off the power of the instrument and place the instrument as shown in Fig 7 5 Take off the filters and clean them Fig 7 5...

Page 34: ...instruments ALMA PARK WIBTOFT LEICESTERSHIRE LE17 5BH UNITED KINGDOM T 44 0 1455 220131 F 44 0 1455 220025 www pginstruments com...

Reviews: