Friabilimeter
operating instructions
Page 18 of 48
Revision 7
© 2016 – Pfeuffer GmbH
Figure 8: Unpacking V
Remove the front hood and the FRIABILIMETER.
Do not lift out the FRIABILIMETER by its drive and
pressure roll.
Note that the weight is approx. 12 kg.
Move the FRIABILIMETER to the installation location, paying attention to the setup instructions in
chapter 5.1.
Keep the original packaging.
Packaging for return delivery
If possible, use the original packaging and the original packaging material. If neither is available any
longer, request new packaging from Pfeuffer GmbH.
5
Installation and commissioning
The Installation and commissioning chapter is only intended for specialist operators.
Installation
5.1
Carefully unpack the FRIABILIMETER (see
chapter 4.4
Place the FRIABILIMETER horizontally on a solid table with a smooth, clean surface.
Make sure there is an adequate distance to all sides so that no heat buildup can occur.
Do not set up the FRIABILIMETER close to apparatus/devices that are sensitive to dust.
Avoid exposure to direct sunlight and extreme ambient conditions.
Make sure that the installation height is ergonomic according to the stature of the operating
personnel.
Connect the supplied mains cable to the FRIABILIMETER using the connector (C19/C20 coupler).
Connect the plug of the mains cable to a suitably earthed socked with protective earth
conductor.
Take the FRIABILIMETER into operation according to the instructions in
chapter 7