Malfunctions
27
9.1 Rectifying malfunctions
Problem
Possible causes
Remedy
Pump will not start up No mains voltage or voltage does
not correspond to the motor data
Check mains voltage and mains fuse protec-
tion; check motor switch
Pump temperature too low
Warm up pump to > 12°C
Thermal protection switch has re-
sponded
Detect and fix cause of overheating; allow
pump to cool off if necessary
Pump system dirty
Clean pump; contact Pfeiffer Vacuum Ser-
vice if necessary
Pump system damaged
Clean and overhaul pump; contact Pfeiffer
Vacuum Service if necessary
Motor defective
Replace motor
Pump switches off af-
ter a while after being
started
Thermal protection switch of the
motor has responded
Detect and fix cause of overheating; allow
motor to cool off if necessary
Mains fuse protection triggered
due to overload (e.g. cold start)
Warm up pump
Exhaust pressure too high
Check opening of exhaust line and exhaust
accessories
Pump does not attain
final pressure
Measurement reading is false
Check gauge, check final pressure without
installation connected
Pump or connected accessories
are dirty
Clean pump and check components for con-
tamination
Operating fluid dirty
Operate pump for a longer period with gas
ballast valve open or change operating fluid
Leak in system
Repair leak
Operating fluid filling level too low Top off operating fluid
Pump damaged
Contact Pfeiffer Vacuum Service
Pumping speed of
pump too low
Intake line not well-dimensioned
Keep connections as short as possible and
see that cross-sections are sufficiently di-
mensioned
Exhaust pressure too high
Check opening of exhaust line and exhaust
accessories
Loss of operating fluid Swivel gasket leaky
Check tightness; replace gasket if neces-
sary
Operational loss of operating fluid If necessary, install oil mist filter and oil re-
turn unit
Unusual operating
noises
Silencer dirty
Clean or replace the silencer;
Damage to the pump system
Clean and overhaul pump; contact Pfeiffer
Vacuum Service if necessary
Motor bearing defective
Replace motor; contact Pfeiffer Vacuum
Service if necessary
NOTICE
Service work should be carried out by qualified personal only!
Pfeiffer Vacuum is not liable for any damage to the pump resulting from work carried out
improperly.
Take advantage of our service training programs; additional information at www.pfeif-
fer-vacuum.com.
Please state all the information on the pump rating plate when ordering spare parts.