Installation
11
5
Installation
5.1
Setting up the pump
When installing the pump, observe the following conditions:
Always place the pump on a firm, even surface.
– Check the load-bearing capacity of the floor at the installation location.
– The vacuum flange serves as the reference surface.
– A drilled hole is provided on each foot for attachment to the mounting surface. Do
not distort the pump when screwing on.
When installing the pump in a closed housing, ensure there is sufficient air circulation.
– Both sight glasses must be visible and accessible for inspection and maintenance
purposes.
– Voltage and frequency information given on the motor rating plate must be visible.
Where stationary installation is involved, anchor the pump on site.
Fill up with lubricant before operating the first time
.
– Amount and type according to rating plate
Installation
conditions
The vacuum pump must be installed and operated under the following ambient condi-
tions:
5.2
Filling with lubricant
The type and amount of lubricant is visible on the pump's rating plate for each vacuum
pump. Generally only the lubricant used during initial assembly can be used later on.
Permissible operating fluids
●
P3 (Standard operating fluid)
●
D1 for special applications (e.g. higher operating temperatures)
– Ultimate pressure measured with gas type dependent gauge: < 5
·
10
-2
hPa
Filling with lubricant
Unscrew lubricant filler screws 15.
Fill gear chamber and bearing chamber (on the drive side) with lubricant.
– Fill level on initial fill: approx. 5 mm above the middle of the sight glass.
Installation location
Inside; must however be protected against dust deposits
Outside; must however be protected against direct weather influ-
ences
Permissible ambient temperature
range
+5 °C ... +40 °C
Relative humidity
Max. 85 %
Installation altitude
Max. 2000 m above m.s.l., the rated power of the motor is re-
duced by approx. 10 % for installation altitudes > 1000 m and an
ambient temperature of 40 °C
Power connections
according to voltage and frequency information given on the mo-
tor rating plate
NOTICE
Use approved operating fluids only!
The use of operating fluids that have not been approved by Pfeiffer Vacuum shall result
in a limited warranty. In such cases, it is not possible to guarantee that product-specific
performance data will be achieved.
Prior consultation is required before using other application-specific operating fluids.