Procedure
1. Utilize the grounding cable on the inside of the electronics housing to discharge applicative inter-
ferences.
2. Connect the eye for M4 grounding cable to a suitable connector.
3. Route the connection in accordance with locally applicable provisions.
5.11 Connect to the mains power supply
WARNING
Risk of fatal injury due to electric shock on account of incorrect installation
The device's power supply uses life-threatening voltages. Unsafe or improper installation can lead to
life-threatening situations from electric shocks obtained from working with or on the unit.
► Ensure safe integration into an emergency off safety circuit.
► Do not carry out your own conversions or modifications on the unit.
2
1
Fig. 18:
Power supply connector and main switch on HiCube Pro
1 Power supply cable
2
Main switch
Establish mains power supply for HiCube Pro
1. Make sure that the master switch is off before connecting.
2. Always ensure a secure connection to the earthed conductor (PE).
3. Plug the power supply cable supplied in the scope of delivery into the power supply plug on the
side of the pumping station housing.
4. Secure the mains cable using a mounting bracket.
5. Connect the power supply cable to the on-site mains power supply.
Installation
36/80