7.2 Maintenance instructions for magnetic coupling
WARNING
Danger of injury from strong magnetic field
There is a risk of injury for people with pacemakers and medical implants.
► Make sure that such individuals do not enter the sphere of influence (≤ 2 m) of the magnetic field.
► Identify rooms in which magnetic couplings are openly accessible with the symbol: "
No access
for people with pacemakers
".
► Always keep disassembled couplings away from computers, data carriers, and other electronic
components.
This safety note applies to
drive system disassembly
in pump versions with magnetic coupling.
7.3 Checklist for inspection and maintenance
Notes on maintenance intervals
Depending on the process, the required maintenance intervals may be shorter than the ref-
erence values specified in the table.
● Consult with Pfeiffer Vacuum Service about shorter maintenance intervals for extreme
loads or for specific processes.
You can carry out maintenance work at
Maintenance Level 1
yourself.
We recommend Pfeiffer Vacuum Service for carrying out maintenance work of
Maintenance Level 2
and
Maintenance Level 3
(revision). If the required intervals listed below are exceeded, or if mainte-
nance work is carried out improperly, no warranty or liability claims are accepted on the part of Pfeiffer
Vacuum. This also applies if original spare parts are not used.
Action
Inspec-
tion
Mainte-
nance
level 1
Mainte-
nance
level 2
Mainte-
nance
level 3
Required
materials
Described in document
OI
OI
MM
MM
Interval
daily
≤ yearly
≤ 2 years
≤ 4
years
Inspection
Visual and acoustic pump check
Check operating fluid:
● Check fill level
● Check color (contamination)
● Check vacuum pump for leaks
■
Check accessories (in accordance
with respective operating instruc-
tions)
■
Maintenance level 1 – operating fluid replacement
Clean the outside of the vacuum
pump:
● Pump housing
● Clean motor fan cap
Changing the operating fluid
■
Operating flu-
id
Maintenance
kit 1
Additional activities:
● Remove cap
● Clean inside of cap and outside
of pumping system (without
cleaning agent)
● Remove and clean gas ballast
valve, replace wearing parts
■
as re-
quired
Maintenance
34/62