Technical data
Units
ATH 500 M
ATH 500 M
ATH 500 MT
Cooling type (acces-
sory)
water/(air)
water/(air)
water
Weight (Air cooling)
Weight (Water cool-
ing)
kg
19
15
19
15
19
15
Backing pump recom-
mended
m
3
/h
15 mini.
15 mini.
15 mini.
1) The products that have customer specificities may have different characteristics.
2) Standard reference conditions: T0 = 273,15 K - P0 = 1013.25 hPa.
3) At nominal speed with 25 °C water cooling temperature.
4) With an exhaust pressure < 0.4 hPa, depends on external conditions (temperature, water flow rate, ambient tempera-
ture). For other conditions, contact us
5) Reduced maximum flow rate, ultimate pressure at exhaust without loss of flow rate speed. Depends on environmental
conditions.
6) With ISO-F flange (best with CF-F flange). Contact us.
7) This maximum temperature influences the maximum flow, contact us.
8) Up to 90 % of full rotation speed, with exhaust pressure < 0.1 hPa.
Tbl. 9:
Technical characteristics
15.2.1 Environmental characteristics
Use
indoor use
Installation altitude
up to 2,000 m
Protection rating (air cooled pump)
Protection rating (water cooled pump)
IP 20
IP 40
Ambient operating temperature (water
cooled pump)
Model M: 5 – 45 °C
Model MT: 5 – 25 °C
Ambient operating temperature (air
cooled pump)
Model M: 5 – 30 °C
Storage temperature
-5 – +50 °C
Maximum relative humidity
80 % max. with T ≤ 31°C, up to 50 % max. with T ≤ 40°C
Pollution degree
2
Transient overvoltage protection
1)
Category II
1) Transient overvoltages up to overvoltage category II levels. Temporary overvoltages that affect the
mains supply.
Tbl. 10:
Environmental characteristics
15.2.2 Cooling water characteristics
The characteristics below concern only the turbomolecular pump with magnetic ball bearings. This
pump is associated with a backing pump. To ensure the effective operation of the application:
► Comply with the most restrictive characteristics for the installation.
► Make sure the water coming from the equipment does not freeze.
Technical data and dimensions
73/78