Pettersson D240x Operating Instructions Manual Download Page 5

In the absence of sound, the detector is continuously recording into the memory. When a sound 
exceeding the chosen trigger level is detected, the detector interrupts the recording and automatically 
replays the stored signal once. A 50% pretrigger function is applied, which means that the detector 
stores the signal starting from a time “50%” prior to the triggering instant. 50% relates to the chosen 
memory size. E.g. with a memory size of 3.4 seconds, the pretrigger time is 1.7 seconds. That way the 
onset of the pulse that triggered the detector, will not be cut off. 
 
The trigger level can be continuously varied with the LEVEL ADJ control. You need a small 
screwdriver or similar to change the LEVEL ADJ control. Turning the control clockwise increases the 
threshold level. Setting the control maximum counter-clockwise will cause the detector to trigger all 
the time, since the noise in the input amplifier then will exceed the threshold level. There is also a 
switch (TRIGGER LEVEL) to enable quic kly changing the threshold. The LOW position will trigger 
the time expansion system at a low signal level, i.e. the detector will capture weaker signals. Changing 
the LEVEL ADJ will affect the sensitivity in both the LOW and HIGH positions. 
 
Normally, all  sounds exceeding the threshold level within the frequency range of the detector will 
trigger the time expansion system. It is, however, also possible to activate the frequency selective 
triggering mode. This is done by setting the TRIGGER SOURCE switch to the HETerodyne position, 
and the output from the heterodyne system will be used as trigger signal instead. That way the setting 
of the heterodyne frequency control (the frequency shown on the display) will determine which 
frequencies the time expansion triggering system will respond to. Setting the heterodyne control to 30 
kHz will capture signals between approximately 25 and 35 kHz. This is useful to reduce the influence 
from background noise or to make the detector capture only certain species. It should  be noted that 
even if the detector has its highest sensitivity near the tuned heterodyne frequency, strong signals 
outside of this range can also trigger the system. In particular, if the detector’s input amplifier is 
overloaded, harmonics will be created, which may cause the triggering system to trigger although the 
fundamental frequency is not within the tuned (heterodyne) frequency range. 
 
 

Heterodyne analysis of the stored signal

 

 
The signal stored in the time expansion memory is normally replayed at one tenth of its original rate. It 
is, however, also possible to play it through the heterodyne system at its original rate. That way the 
user can analyze the stored sequence carefully, e.g. checking for max and min frequencies etc. This 
mode is also useful to demonstrate the heterodyne principle in front of an audience. 
 
To feed the stored signal through the heterodyne system, the NORMAL/ TE

HET switch should 

be in the TE

HET position and the HET/TIME EXP switch in the HET position. 

 
This mode is intended for use in the manual triggering mode. If you wish to use it in the automatic 
triggering mode, you should change the TRIG switch to MAN immediately after the signal you want 
to study has been captured, or the detector will start recording again after replaying the sequence once. 
 
 

Using the relay output 

 

In some situations, you may want to automatically dump the time expanded calls onto a tape recorder. 
This can be achieved using a voice activated tape recorder. However, the D240x also has an output 
which can be used to control a relay, which in turn can turn on/off a tape recorder. A separate relay 
unit will be available from Pettersson Elektronik.  
 
The relay control output should be used only in the automatic triggering mode. It will be activated 
when a signal exceeds the triggering threshold level, and remains activated while the sequence is 
replayed once.  
 

Reviews: