5. If the scale displays [O_Ld] the load exceeds the maximum capacity. 

Please reduce the weight immediately. 

 

 
V. Tare 

1. Press the 

ON/OFF

 button and place a bowl or any another container on 

the platform. Its weight should not exceed the maximum capacity of the 
scale. 

2. Press 

TARE

 and wait until [0:000] is displayed. 

3. Add the load and directly read its net weight. 
 
 

VI. Counting Function

 

1. Press 

ON/OFF

 (possibly with an adequate tare bowl on the platform) and 

wait until [0:000] is displayed or press 

TARE

 to get the [0:000] reading.  

2. Press and hold the 

PCS

 key for 3 seconds. Any of the four possible 

reference lots [25], [50], [100] or [200] e.g. [50] will be displayed. 

3. If applicable press 

MODE

 to select another reference lot e.g. [100]. 

Info:

 

the higher the lot quantity the better the precision of the following count.  

4. Place the selected reference lot (25, 50, 100 or 200 pcs) of the identical 

items to be counted onto the platform or into the tared container.  

5. Briefly press the 

PCS

 key to confirm the selected reference lot. [PCS] will 

appear on the display to confirm the calibration of the counting function. 
The scale is now in counting mode. 

6. Add more identical items and their quantity will be displayed (not their 

weight). 

7. If required switch to weighing mode by pressing 

MODE

 and back to 

counting mode by pressing 

PCS

Info:

 the last reference lot selected as 

well as its weight will remain stored even if the scale is switched off. 

 
 

VII. Professional 2 step calibration (only in gram unit) 

1. Press 

ON/OFF

 and give the scale 30 seconds to warm up. 

2. Wait until [0.0] is displayed. Press and hold the 

TARE

 key for 3 

– 5 

seconds. 

3. As soon as [CAL] is displayed, release the 

TARE

 key and the first required 

calibration weight will be displayed with flashing [1000.0]. 

4. Gently add a 1000g calibration weight to the centre of the platform. The 

display will change to flashing [2000.0] requiring the second step 
calibration weight. 

5. Add another 1000g calibration weight or remove the first one and gently 

add a 2000g calibration weight. The display will switch to [PASS] and 
return to normal weighing mode, still displaying [2000.0].  