1.2 Environmental Requirements
To insure correct operation and stable performance over an extended period of time, install the T6U in
a location that meets the following conditions:
1.
Operation conditions:
Condition
Altitude
Temp (º
C)
Humidity (%RH)
Non-condensing
Non-operating
(Transport)
0-2000m
5-45
20-80
Non-operating
&meeting
dielectric
strength tests
Sea level
40
91-95
Operating but
not
necessarily
meeting
performance
specifications
0-2000m
5-31
31-40
≤ 80
≤ [80-3.33(t-31)]
Operating
within
performance
specifications
0-853m
853-2000m
10-35
10-25
8-80
8-80
2. Protection degree: IP20
3.
Avoid magnetic fields
Away from the electric appliances with magnetic field, electromagnetic field and high frequency
waves;
4. Avoid corrosive gas
Area free from corrosive gas, such as chlorine, hydrochloric acid gas, hydrogen sulphides gas,
sulfurous acid gas;
5.
Area with very little dust or other airborne particles.
6.
Placing on the stable workbench
The workbench should be stable and strong enough to support the total weight of equipment to be used.
The worktops should be large enough to permit a free circulation of air around each unit in the system.
Remember to provide space for the computer, monitor and printer. The workbench should be about 90cm
high. To avoid damage through spillage of the samples being analyzed, the worktops should be covered
with a material that is corrosion resistant and impervious to liquids.
7.
Do not share a power socket with other electric appliances.
Equip the instrument with a power socket, which should not be shared with other electric appliance.
The power should be with protective ground wire.
8.
Avoid direct sunlight
The improper environment will affect the measurement result and the
instrument service life!