ROHO nexus SPIRIT Cushion Operation Manual
Cleaning and Disinfecting
Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. Before
use by a different individual: clean, disinfect, and check product for proper functioning.
Cover Care: Before cleaning and disinfecting, remove the cover from the cushion. Remove the hook fastener
pieces from bottom of the cover.
To Clean the Cover: Machine wash in cold water (30°C) with mild detergent,
gentle cycle, or wipe clean with neutral detergent and cold water. Tumble dry low.
To Disinfect the Cover: Hand wash in cold water (30°C) using 1 part household
liquid bleach per 9 parts water, then rinse thoroughly. Or machine wash in
hot water (60°C). Tumble dry low.
Follow the safety guidelines on the bleach container.
Cushion Care Cautions: Valve(s) must be closed. • DO NOT allow water or cleaning solution to enter
the cushion insert. • DO NOT machine wash or machine dry the cushion insert or the contoured foam base. •
Use of the following may cause damage to the cushion: abrasives (steel wool, scouring pads); caustic,
automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents, including
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based lotions,
lanolin, or phenolic-based disinfectants; ozone gas; ultraviolet light cleaning methods.
To Clean the Cushion Insert: Carefully remove the cushion insert from the contoured foam base. Close the
inflation valve. Hand wash, using mild liquid hand soap, hand dishwashing liquid, laundry detergent, or a
multipurpose detergent (Follow product label instructions.) Gently scrub all cushion surfaces, using a soft
plastic bristle brush, a sponge or a cloth. Rinse with clean water. Air dry. DO NOT expose to direct sunlight.
To Disinfect the Cushion Insert: Repeat cleaning instructions, using 1 part household liquid bleach per 9
parts water. Follow safety guidelines on bleach product label. Keep the product wet with the bleach solution
for 10 minutes. Rinse thoroughly with clean water. Air dry.
To Clean the Contoured Foam Base: Remove the cover. Wipe gently with a cloth dampened with
water and a household detergent. Rinse cloth and re-wipe.
Allow foam to air dry thoroughly before putting the cover on.
This symbol means, “Wipe with a damp cloth.”
Disinfectants are not effective on porous surfaces such as foam. If the foam portion of the
product becomes soiled or contaminated, DO NOT use with multiple individuals.
Note: Most germicidal disinfectants are safe if used following the disinfectant manufacturer’s dilution direc-
tions.
Note About Sterilization: High temperatures accelerate aging and will damage the product assembly.
ROHO products are not packaged in a sterile condition, nor are they intended or required to be sterilized prior
to use. If institution protocol demands sterilization: First, follow the cleaning and disinfecting instructions;
then, open the inflation valve and use the lowest possible sterilizing temperature, but not to exceed 70°C
(158°F), for the shortest time possible. DO NOT autoclave.
30C
30C
30C
1:9
+