background image

6

ROHO nexus SPIRIT Cushion Operation Manual

Cleaning and Disinfecting 

Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. Before 
use by a different individual: clean, disinfect, and check product for proper functioning.

Cover Care: Before cleaning and disinfecting, remove the cover from the cushion. Remove the hook fastener 
pieces from bottom of the cover.
To Clean the Cover: Machine wash in cold water (30°C) with mild detergent, 
gentle cycle, or wipe clean with neutral detergent and cold water. Tumble dry low.

To Disinfect the Cover: Hand wash in cold water (30°C) using 1 part household 
liquid bleach per 9 parts water, then rinse thoroughly. Or machine wash in 
hot water (60°C). Tumble dry low.

Follow the safety guidelines on the bleach container.

Cushion Care Cautions: Valve(s) must be closed. • DO NOT allow water or cleaning solution to enter 

the cushion insert. • DO NOT machine wash or machine dry the cushion insert or the contoured foam base. • 
Use of the following may cause damage to the cushion: abrasives (steel wool, scouring pads); caustic, 
automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents, including 
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based lotions, 
lanolin, or phenolic-based disinfectants; ozone gas; ultraviolet light cleaning methods.

To Clean the Cushion Insert: Carefully remove the cushion insert from the contoured foam base. Close the 
inflation valve. Hand wash, using mild liquid hand soap, hand dishwashing liquid, laundry detergent, or a 
multipurpose detergent (Follow product label instructions.) Gently scrub all cushion surfaces, using a soft 
plastic bristle brush, a sponge or a cloth. Rinse with clean water. Air dry. DO NOT expose to direct sunlight.
To Disinfect the Cushion Insert: Repeat cleaning instructions, using 1 part household liquid bleach per 9 
parts water. Follow safety guidelines on bleach product label. Keep the product wet with the bleach solution 
for 10 minutes. Rinse thoroughly with clean water. Air dry. 
To Clean the Contoured Foam Base: Remove the cover. Wipe gently with a cloth dampened with 
water and a household detergent. Rinse cloth and re-wipe. 
Allow foam to air dry thoroughly before putting the cover on.

This symbol means, “Wipe with a damp cloth.”

Disinfectants are not effective on porous surfaces such as foam. If the foam portion of the 

product becomes soiled or contaminated, DO NOT use with multiple individuals.

Note: Most germicidal disinfectants are safe if used following the disinfectant manufacturer’s dilution direc-
tions.

Note About Sterilization: High temperatures accelerate aging and will damage the product assembly. 

ROHO products are not packaged in a sterile condition, nor are they intended or required to be sterilized prior 
to use. If institution protocol demands sterilization: First, follow the cleaning and disinfecting instructions; 
then, open the inflation valve and use the lowest possible sterilizing temperature, but not to exceed 70°C 
(158°F), for the shortest time possible. DO NOT autoclave.

30C

30C

30C

1:9

+

Summary of Contents for ROHO nexus SPIRIT Cushion

Page 1: ...oduct read the instructions and save for future reference Supplier this manual must be given to the user of this product Operator individual or caregiver before using this product read the instructions and save for future reference s h a p e f i t t i n g t e c h n o l o g y ROHO nexus SPIRIT Cushion Operation Manual ...

Page 2: ......

Page 3: ...leville IL 62221 5429 U S A U S A 800 851 3449 Fax 888 551 3449 Customer Care orders roho permobil com Outside the U S A 618 277 9150 Fax 618 277 6518 intlorders roho permobil com permobilus com ROHO products may be covered by one or more U S and foreign patents and trademarks including ROHO DRY FLOATATION shape fitting technology and nexus SPIRIT ROHO Inc has a policy of continual product improve...

Page 4: ...when com pared to normal skin may indicate the beginning of soft tissue breakdown and may need to be addressed If there is any discoloration to skin soft tissue STOP USE immediately If the discoloration does not disappear within 30 minutes after disuse immediately consult a healthcare professional The cushion and the cover MUST be compatible sizes and MUST be used as directed in this manual IF THE...

Page 5: ...t neoprene rubber made without latex Inflation Valve nickel plated brass Contoured Foam Base polyurethane Cover two way stretch fluid resistant top and sides with a zipper located in the back of the cover and a non skid bottom with hook and loop fasteners to help secure the cushion in the chair Refer to the cover care label for material content Cushion Width in Depth in Width cm Depth cm NS1414C 1...

Page 6: ...ir keeping your hand beneath the individual s lowest bony prominence Release air until you are barely able to move your fingertips Close the valve Note For two valve cushions repeat all steps for both valves Important Safety Information Check inflation frequently at least once a day DO NOT use an under inflated cushion or an over inflated cushion because the product benefits will be reduced or eli...

Page 7: ...ation valve s are open Inspect the inflation valve s and hose s for damage Uncomfortable or Unstable Confirm that the cushion insert is not over inflated See Cushion Adjustment the air cells and the top of the cover are facing up the cushion is centered beneath the individual the cushion is properly sized to the individual and the wheelchair Allow the individual at least one hour to adjust to the ...

Page 8: ...oured foam base Close the inflation valve Hand wash using mild liquid hand soap hand dishwashing liquid laundry detergent or a multipurpose detergent Follow product label instructions Gently scrub all cushion surfaces using a soft plastic bristle brush a sponge or a cloth Rinse with clean water Air dry DO NOT expose to direct sunlight To Disinfect the Cushion Insert Repeat cleaning instructions us...

Page 9: ......

Page 10: ...ling Shipping Address 1501 South 74th Street Belleville IL 62223 5900 USA Administrative Offices 100 N Florida Avenue Belleville IL 62221 5429 USA U S 800 851 3449 Fax 888 551 3449 Outside the U S 618 277 9150 Fax 618 277 6518 permobilus com ...

Reviews: